Tilt Actuator and Micro-Switch Removal

The following procedures describe the removal of the tilt actuator and the micro-switch assembly used
for the creep speed. Read the directions entirely prior to performing the task. Portions of the directions
may be eliminated if only the tilt actuator needs to be replaced, or the micro-switch replaced separately.
Reverse the procedures to install the tilt actuator or micro-switch and retest as necessary.
1.
Disconnect the 6-pin connector located on
the rear of the tilt system (fi g. 7.38).
2.
Remove the cover over the 6 pin connector
on the rear of the tilt system using a 3 mm Allen
wrench (fi g. 7.39).
3.
Locate the micro-switch assembly and
remove using a 3 mm Allen wrench (fi g. 7.40).
Note: this step may be eliminated if the micro-
switch is to be disconnected and left in place.
SUNRISE MEDICAL
Section 7
Removal Procedures (cont.)
PULSE 6 SERVICE MANUAL
2009
PAGE 7.13
fig. 7.38
fig. 7.39
fig. 7.40