  • Page 1 User instructions boso ABI-system 100 0124...
  • Page 2: Table Of Contents

    9 Installing the ABI system 100 software • Installing the single user version • Installing the network version 10 Installing the USB connection cable 11 Using the ABI-system 100 software • Starting the program • Calling up the Help function • Configuration before first use •...
  • Page 3 18 Technical data 19 Instructions for disposal 20 Calibration checks - Testing instructions • Function testing • Testing of pressure circuit integrity and deviation of pressure display • Safety seal 21 EMC notes for your boso ABI-system 100 22 Your notes...
  • Page 4: Contents Of Package

    1 Contents of package ABI-system 100 CA04 upper arm cuffs CL04 ankle cuffs ABI-system 100 software USB cable boso power supply unit art. no.: 410-7-154 Warranty certificate User instructions 2 Minimum system requirements • Pentium II or higher • 256 MB RAM •...
  • Page 5: Explanation Of Symbols And Operating Controls

    3 Explanation of symbols and operating controls Cuff connection Cuff connection Right arm Left arm (colour code red) (colour code yellow) Cuff connection Cuff connection Right leg Left leg (colour code black) (colour code green) START key Press this key to start an ABI measurement manually. All 4 cuffs will be inflated.
  • Page 6: Explanation Of Icons

    4 Explanation of icons Read the user instructions Protection class II device Degree of protection against electric shock: type BF USB 2.0 connection The CE mark documents compliance with the Medical Devices 0124 Directive 93/42/EEC, Notified body: DEKRA, 0124 Year of manufacture Manufacturer Potential equalisation Do not dispose of this device in the domestic waste.
  • Page 7: Preliminary Remarks

    5 Preliminary remarks Your boso ABI-system 100 is an innovation on the market of blood pressure mea- surement technology. It provides a simple way of determining the ankle-brachial index (ABI) and works on the oscillometric measurement principle. The pressure variations (oscillations) caused by pulse waves and transmitted by the cuffs are stored and evaluated by microprocessors.
  • Page 8: Ankle-Brachial Index (Abi)

    The mains connection socket is located at the rear of the device. Use only the boso power supply unit (order no. 410-7-154), as it produces a rectified output of the correct polarity. Other commercially available power supply units can cause damage to the electronic components, which will invalidate the warranty.
  • Page 9: Installing The Single User Version

    Installing the single user version Click the [Installiere ABI-Software] button under "Einzelplatzversion" and follow the dialogue on the screen. Installing the network version 1. Install the Firebird-DB-server locally on the server computer by clicking the [Installiere Firebird-DB-Server] button (this step is not required if a Firebird server is already running on the system).
  • Page 10: Installing The Usb Connection Cable

    Press the F1 key or the [Help] button at any time to call up these user instruc- tions as a PDF file. Configuration before first use Before starting the boso ABI system 100, you can enter the following settings ([Settings...] button).
  • Page 11: Entering Doctor's Data

    Windows directory structure directly. boso import file = export file from the practice EDP system boso export file = import file to the practice EDP system Entering limits for the ABI assessment The "ABI-calssification" tab...
  • Page 12: Definition Of Output Format

    Definition of output format Use the “Printer settings” tab to define whether the results are output in paper format, as print preview and/or as a PDF document by clicking on the appropriate [Print...] button during the program run. If "Printer” is selected, click the [Print...] button during the program run to display the Output Options"...
  • Page 13: Patient Tab

    Patient tab On starting the program you are taken to the ABI patient manage- ment screen. If the patient con- cerned is already registered, click on the listing to select them. If the program was started from the practice system, the patient will be automatically selected or created by GDT import.
  • Page 14: Measurement Tab

    Measurement tab Previous measurements can be viewed and new measurements taken in the "Measurements” tab. To display previous ABI measure- ments, select the required ABI measurement in the “Date-Time” field. The following parameters will be displayed on screen (for the left and right side of the body, respectively): systolic blood pressure of the upper arm, displayed in red if >140 mmHg diastolic blood pressure of the upper arm, displayed in red if >...
  • Page 15: Profiles Tab

    Click on one of the oscillation profiles to enlarge the selected profile. The “Pulse level / time” tab shows the oscillati- on curve over the time axis. The “Pulse level / cuff pressure” tab shows the oscillation curve over the cuff pressure. Profiles tab The ABI values of all measurements are displayed gra-...
  • Page 16: Preparing For An Abi Measurement

    The patient should rest for about 5 minutes before a measurement is taken. Attaching the cuff The device should only be used with original boso CA04 or CL04 cuffs Type Circumference Order no.
  • Page 17: Taking An Abi Measurement

    Select “Start new measurement” in the “Measurements” tab or press the START button on the apparatus. The ABI-system 100 starts the ABI measurement after a short calibration process (about 3 seconds). The apparatus has an intelligent automatic inflation system to ensure comfortable inflation to the correct cuff pressure.
  • Page 18: Saving An Abi Measurement

    Saving an ABI measurement To save an ABI measurement, click on the [Save measurement] button. Repeating an ABI measurement Allow an interval or at least 2 minutes between successive measurements. To repeat an ABI measurement press the [Start new measurement] button. To take a measurement on one limb only, press the START key next to the corre- sponding symbol.
  • Page 19: Cleaning The Device And Cuff

    15 Cleaning the apparatus and cuff Please use only a soft, dry cloth to clean the device. Small stains on the cuff can be carefully removed with a damp cloth. 16 Disinfecting the apparatus We recommend Antifect® Liquid disinfectant (Schülke&Mayr) to wipe the appara- tus and cuffs for disinfection.
  • Page 20: Technical Data

    18 Technical data Measurement principle: oscillometric Measurement range: 40 to 240 mmHg Cuff pressure: 0 to 300 mmHg Maximum deviation of cuff ± 3 mmHg pressure measurement: Operating conditions: Environmental temperature 10 to 40°C Relative humidity 10 to 85% Transport and storage conditions: Environmental temperature -5°C to 50°C Relative humidity max.
  • Page 21: Calibration Checks - Testing Instructions

    “Test mode" and click on the “Start” button. After a short calibration process the apparatus will be in test mode. The current pressure will be displayed in the corresponding ABI-system 100 software fields. Perform the usual test for deviation of pressure display and pressure cir cuit integrity (allow a set-up time of at least 30 seconds for the cuff).
  • Page 22: Emc Notes For Your Boso Abi-System 100

    Guidance and manufacturer’s declaration – electromagnetic emissions The boso unit is intended for use in the electromagnetic environment specified below. The customer of the user of the boso unit should assure that it is used in such an environment.
  • Page 23 Guidance and manufacturer’s declaration – electromagnetic immunity The boso unit is intended for use in the electromagnetic environment specified below. The customer or the user of the boso unit should assure that it is used in such an environment. Immunity...
  • Page 24: Your Notes

    22 Your notes...
  • Page 25 Your notes...
  • Page 26 Your notes...
  • Page 27 Your notes...
  • Page 28 BOSCH + SOHN GmbH u. Co. KG Bahnhofstraße 64 • 72417 Jungingen • Germany Telefon: +49 (74 77) 92 75-0 • Fax: +49 (74 77) 10 21 Internet: www.boso.de • e-Mail: zentrale@boso.de...

Table of Contents