Cleaning and Disinfecting Procedures

NOTE
- Clean and disinfect the outside of the device only. Do not remove battery cover when cleaning and disinfecting.
To clean the meter:
1. Throughly wipe the entire surface of the meter with disinfecting wipes listed above to
clean any possible dirt, dust, blood and other body fluids.
To disinfect the meter:
2. Take another disinfecting wipe and wipe the meter thoroughly. (Note: All blood and
body fluids should be cleaned from surface before performing the disinfecting
procedure)
3. Allow the surface to remain wet for 2 minute.
4. Allow to air dry.
NOTE
- Your GE200 Blood Glucose Meter has been tested to ensure that there is no change in the performance or
external materials of the device after 550 cleaning cycles and 550 disinfecting cycles. The testing simulates
2 cleaning and disinfecting cycles per week over the typical life of the meter (5 years).
48
To clean the lancing device:
1. Throughly wipe the entire surface of the lancing device with disinfecting wipes
listed above to clean any possible dirt, dust, blood and other body fluids.
1
2
To disinfect the lancing device:
2. Take another disinfecting wipe and wipe the Lancing device thoroughly. (Note: All
blood and body fluids should be cleaned from surface before performing
disinfecting procedure)
3
3. Allow the surface to remain wet for 2 minute.
4. Allow to air dry.
Allow to remain
wet for 2 minute
NOTE
- Your GE Lancing Device has been tested to ensure that there is no change in the performance of the
Cleaning and Disinfecting Procedures
device after550 cleaning cycles and 550 disinfecting cycles. The testing simulates 2 cleaning and
disinfecting cycles per week over the typical life of the meter (5 years).
1
2
3
Allow to remain
wet for 2 minute