Cleaning:
statMAP
or other liquid. Do not use alcohol.
Do not use steam or heat to sterilize the statMAP.
Cuffs, Tubing and Port Fittings
cuff, remove it from the statMAP, close off the tube with a plug,
and place the hook and loop fasteners in closed position.
Use one or more of the following methods and allow to dry:

Wipe with mild detergent and water solution (1:9 solution).
Rinse.

Wipe with Enzol per manufacturer's instructions. Rinse.

Wipe with 0.5% bleach and water solution. Rinse.

Wipe with 70% isopropyl alcohol.
Calibration:
Calibration of the statMAP should be checked annually or
when there is doubt about the validity of the measurements.
Proper calibration is indicated if needle and LED are in
alignment during pump up AND subsequent deflation.

MAINTENANCE

: Clean with damp cloth. Do not immerse in water