OPERATING INSTRUCTIONS:
Semi-Automatic Mode (con't)
11. When both AMBR lights flash in the status indicator areas,
the BP determination is complete. The remaining air must
be released from the cuff by squeezing the deflate trigger.
The statMAP will indicate BP and Heart Rate at the end of
the deflation on the LEDs to the nearest 5mmHg tic mark.
To obtain a BP reading to the nearest 1mmHg, refer to
page 21.
Solid LED Illuminations
Highest reading - Systolic
Lowest reading - Diastolic
Diastolic
Heart Rate
)
(Flashing
12. To obtain another reading, just re-inflate the cuff. statMAP
"remembers" the last 10 readings in a session and uses
them to compute the NSV.
NOTE: statMAP remembers all readings but can only
generate a good NSV if there are 3 or more readings
made in a session.
13. The NSV is displayed for the user by cycling the power
OFF then ON after a BP measurement session is finished.
All of the session readings are then analyzed by the
statMAP and a "Nominal Session BP Value" is displayed.
Flashing LED Illumination
Heart Rate
Systolic
Page 19
NOTE:
If there is only 1
solid LED &
1 flashing LED,
the flashing LED
represents the
HR, AND either
the Systolic or
the Diastolic
pressure.