Ramsey Medical statMAP Instructions Manual page 10

Blood pressure measurement device
Table of Contents

Advertisement

Page 10
The statMAP is intended for use in the electromagnetic
environment specified below. The customer or user of the
statMAP should assure that it is used in such an environment.
Immunity test
Electrostatic
discharge
(ESD)
IEC 61000-4-2
Electrical fast
transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips,
short
interruptions
and voltage
variations on
power supply
input lines
IEC 61000-4-11
Power
frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
Electromagnetic Immunity
IEC60601
test level
±6 kV
contact
± 8 kV air
±2 kV for
power supply
lines
±1 kv for
input/output
lines
± 1 kV
differential
mode
± 2 kV
common
mode
>95% dip in
0.5 cycle
60% dip in 5
cycles
30% dip for
25 cycles
>95% dip for
5 seconds
3 A/m
Compliance
Electromagnetic
Level
environment - guidance
Floors should be wood,
concrete or ceramic tile.
±6 kV
If floors are covered with
contact
synthetic material, the
± 8 kV air
relative humidity should
be at least 30%.
not
applicable
not
applicable
not
applicable
not
applicable
not
applicable
not
applicable
not
applicable
not
applicable