Recommended Equipment; Environmental Conditions; Warm Up - ILX Lightwave Temperature Controller LDT-5525 User Manual

Hide thumbs Also See for Temperature Controller LDT-5525:
Table of Contents

Advertisement

3
C H A P T E R

Recommended Equipment

Recommended test equipment for calibrating the LDT-5525 Temperature
Controller is listed in Table 3.1. Equipment other than that shown in the table may
be used if the specifications meet or exceed those listed.
Table 3.1
DESCRIPTION
DMM
Resistors

Environmental Conditions

Calibrate this instrument under laboratory conditions. We recommend calibration
at 23
Controller may be calibrated at its intended use temperature if this is within the
specified operating temperature range of 0 to 40

Warm Up

The LDT-5525 Temperature Controller should be allowed to warm up for at least 1
hour before calibration.
20
LDT-5525
M A I N T E N A N C E A N D T R O U B L E S H O O T I N G
Calibration Overview
MFG / MODEL
HP 3457A
Metal Film
High Power
o
o
C ± 1.0
C. When necessary, however, the LDT-5525 Temperature
SPECIFICATION
DC Amps (@ 1.0A): +0.02%
Resistance (@ 10Ω): 0.02%
0.1 µA or 0.1 mW resolution
15 KΩ (for ITE calibration)
4 KΩ and 40 KΩ (for 100 mA calibration)
4 KΩ and 400 KΩ (for 10 mA calibration)
4 KΩ and 10 KΩ (for LM335 calibration)
10 KΩ and 20 KΩ (for AD590 calibration)
1 Ω, 20W, low TCR (for ITE calibration)
o
C.