ILX Lightwave Temperature Controller LDT-5525 User Manual page 40

Hide thumbs Also See for Temperature Controller LDT-5525:
Table of Contents

Advertisement

3
C H A P T E R
Unable to adjust output or
parameter
Unable to switch DISPLAY,
MODE, SENSOR CAL or
PARAMETER modes
Output goes off intermittently
R Mode set point is not saved
26
LDT-5525
M A I N T E N A N C E A N D T R O U B L E S H O O T I N G
Troubleshooting
Check the ADJUST ENBL switch; the indicator must be lit for any
adjustments to be made
Check the MODE or DISPLAY switch; if they do not respond, the
unit may be in measurement calibration mode (see Chapter 3);
press the OUTPUT switch to abort this mode.
The unit may be in measurement calibration mode (see Chapter
3); press the OUTPUT switch to abort this mode
Check that the AC power cord connection is secure; power line
drop outs may reset the unit and when the power is restored, the
output will be off
Check the TE module connections; a high impendance on the TE
load may cause the output to exceed the compliance voltage
momentarily, thus shutting the output off
The R mode set point value is not independent from the T mode
set point value; if the control mode is changed from R mode to T
mode, the R set value will change to a value which corresponds to
the temperature, based on C1 and C2