EPS Bio Technology EasyMax Voice III User Manual page 49

Self-monitoring blood glucose system
Hide thumbs Also See for EasyMax Voice III:
Table of Contents

Advertisement

10. Critically ill patients should not be tested with home-use blood glucose meters.
11. Interfering Substances depend on the concentration. The below substances up to the test concentration
will not affect the test results.
Concentrations
of the interference tested
Ascorbic Acid,3 mg/dL (0.17mmol/L)
Ibuprofen,40 mg/dL (1.94mmol/L)
L-Dopa,10 mg/dL (0.51mmol/L)
Sodium Salicylate,50 mg/dL (3.12mmol/L)
Tetracycline,0.4 mg/dL (0.01mmol/L)
Tolbutamide,100 mg/dL (3.70mmol/L)
Bilirubin-unconjugated,1.2 mg/dL (0.02mmol/L)
Uric acid,14 mg/dL (0.83mmol/L)
Xylose,20 mg/dL (1.33mmol/L)
Glucose
Bias
Level
50 mg/dL
250 mg/dL
(2.8 mmol/L)
(13.9 mmol/L)
9.5
-2.0
85.1
1.8
0.4
0.8
-0.2
26.8
1.4
500 mg/dL
(27.8 mmol/L)
6.0%
1.0%
-1.8%
-1.7%
30.5%
1.1%
-0.6%
-0.1%
2.1%
-0.6%
-1.5%
1.0%
-3.7%
-5.5%
5.6%
9.3%
4.2%
5.5%