Summary of Contents for Clever Choice Auto-Code Pro
Page 1
Distributed by Simple Diagnostics P.O. Box 128, Williston Park, NY 11596 USA 24-HOUR TOLL-FREE TELEPHONE ASSISTANCE For in vitro diagnostic use only 1-877-DIABETK 1-877-342-2385 Read instructions before use www.SimpleDiagnostics.com For self-testing...
Blood Glucose Monitoring System OWNER'S MANUAL Dec. 2009 Version 1.0 311-4267100-001...
Page 4
Regular monitoring of your blood glucose levels can help you and your doctor gain better control of your diabetes. Due to its compact size and easy operation, you can use the CLEVER CHOICE Auto- Code Pro Blood Glucose Monitoring System to easily monitor your blood glucose levels by yourself anywhere, any time.
IMPORTANT SAFETY PRECAUTIONS READ BEFORE USE 1. Use this device ONLY for the intended use described in this manual. 2. Do NOT use accessories which are not specified by the manufac- turer. 3. Do NOT use the device if it is not working properly or or if it is damaged.
TABLE OF CONTENTS IMPORTANT SAFETY PRECAUTIONS BEFORE YOU BEGIN Important Information Intended Use Test Principle Contents of System Meter Overview Screen Display Test Strip Setting the Meter THE FOUR MEASURING MODES CONTROL SOLUTION TESTING Performing a Control Solution Test TESTING WITH BLOOD SAMPLE Overview of the Lancing Device Setting up the lancing device Preparing the puncture site...
Page 7
MAINTENANCE Battery Caring for Your Meter Caring for Your Test Strips Important Control Solution Information SYSTEM TROUBLESHOOTING Result Readings Error Messages Troubleshooting DETAILED INFORMATION Reference values Comparing Meter and Laboratory Results SPECIFICATIONS...
BEFORE YOU BEGIN Important Information Severe dehydration and excessive water loss may cause readings which are lower than actual values. If you believe you are suffering from severe dehydration, consult a healthcare professional im- mediately. If your blood glucose results are lower or higher than usual, and you do not have symptoms of illness, first repeat the test.
Intended Use This system is intended for use outside the body (in vitro diagnostic use) by people with diabetes at home and by health care profes- sionals in clinical settings as an aid to monitoring the effectiveness of diabetes control. It is intended to be used for the quantitative measurement of glucose (sugar) in fresh whole blood samples (from the finger, palm, forearm, upper arm, calf and thigh).
Contents of System Your new CLEVER CHOICE Auto-Code Pro system kit includes: Daily Log Book Meter Owner’s Manual Warranty Card Storage case Sterile Lancets Quick Start User Guide One CR2032 Battery Quick Start User Guide Test strip and control solution are not included in the kit (please check the contents on your product box).
Meter Overview Left Side Front Right Side Back Rear DISPLAY SCREEN TEST SLOT Insert test strip here to turn the meter BATTERY COMPARTMENT on for testing. M BUTTON DATA PORT Enter the meter memory and silence a reminder alarm. Download test results with a cable connection.
Screen Display Measurement mode Day average Blood drop symbol Low battery symbol Test strip symbol Memory symbol Error message Ketone warning Test result Control solution mode Date Measurement unit Time Reminder alarms...
The front side of the test strip should face up when inserting the test strip. Front Back NOTE The CLEVER CHOICE Auto-Code Pro meter should only be used with CLEVER CHOICE Auto-Code Pro Test Strips. Using other test strips with this meter can produce inaccurate results.
Setting the Meter Before using your meter for the first time or if you change the meter battery, you should check and update these settings. Make sure you complete the steps below and have your desired settings saved. Entering the setting mode Start with the meter off (no test strip inserted).
Page 15
2. Setting the time format Press and release to select the desired time format --- 12h or 24h. Press 3. Setting the time With the hour flashing, press until the cor- rect hour appears. Press With the minute flashing, press until the cor- rect minute appears.
Page 16
If you would like to delete ALL the results, press and select “Yes” then press to confirm. “OK” and “ ” are displayed on the meter, which indicates that all data stored is deleted. 5. Setting the Reminder Alarm You may set up any or all of the reminder alarms (1-4). The meter displays “On”...
Page 17
If you do not want to set an alarm, press to skip this step. If you want to turn off an alarm, find the alarm number by pressing in the setting mode, press to change from “ON’ to “OFF”. At the time of your alarm, the meter will beep and automatically turn on.
THE FOUR MEASURING MODES The meter provides you with four modes for measuring General, AC, PC and QC. Modes Use when any time of day without regard to time since General (displays as “Gen”) last meal no food intake for at least 8 hours 2 hours after a meal testing with the control solution You can switch between each mode by:...
CONTROL SOLUTION TESTING CLEVER CHOICE Control Solution contains a known amount of glucose that reacts with test strips and is used to ensure your meter and test strips are working together properly. Do a control solution test when: you first receive the meter,...
Page 20
2. Press to mark this test as a control solu- tion test With “QC” displays, the meter will not store your test result in memory. If you press again, the “QC” will disappear and this test is no longer a control solution test.
Page 21
4. Read and Compare the Result After counting down to 0, the test result of control solution will appear on the display. Compare this result with the range printed on the test strip vial and it should fall within this range. If not, please read instructions again and repeat the control solution test.
TESTING WITH BLOOD SAMPLE Overview of the Lancing Device Lancet Holder Cocking Control Release Button Depth Indicator A Clear Cap Arrow Adjustable Tip If your lancing device differs from the one shown here, please refer to the manufacturer’s manual to ensure proper usage. NOTE To reduce the chance of infection: Never share a lancet or the lancing device.
Setting Up the Lancing Device Screw off the cap of the lancing device. Insert a lancet into the lancet holder and push down firmly until it is fully in place. Twist the protective disk until it separates from the lancet. Replace the lancing device cap.
Preparing the Puncture Site Stimulating blood perfusion by rubbing the puncture site prior to blood extraction has a significant influence on the glucose value obtained. Blood from a site that has not been rubbed exhibits a measurably different glucose concentration than blood from the finger. When the puncture site was rubbed prior to blood extraction, the difference was significantly reduced.
Fingertip testing Hold the lancing device firmly against the side of your finger. Press the release button. You will hear a click, indicating that the puncture is complete. Blood from sites other than the fingertip A clear cap, together with the kit, makes it easier to get a drop of blood for AST.
Alternative Site Testing IMPORTANT: There are limitations with AST (Alternative Site Testing). Please consult your health care professional before you per- form AST. What is AST? Alternative site testing (AST) means that people use parts of the body other than the fingertips to check their blood glucose levels.
Page 27
We strongly recommend you perform AST ONLY at the following times: In a pre-meal or fasting state (more than 2 hours since the last meal). Two hours or more after taking insulin. Two hours or more after exercise. Do NOT use AST if: You think your blood glucose is low.
Performing a Blood Glucose Test 1. Insert the test strip to turn on the meter Wait for the meter to display the test strip and blood drop symbol. 2. Select the appropriate measurement mode by pressing For selecting the measurement mode of AC, PC, or Gen, please refer to the “THE FOUR MEASURING MODES”...
Page 29
NOTE Do not press your puncture site against the test strip or try to smear the blood. If you do not apply a blood sample to the test strip within 3 minutes, the meter will automatically turn off. You must remove and reinsert the test strip to start a new test.
Page 30
6. Eject the used test strip and remove the lancet To eject the test strip, point the strip at a dis- posal container for sharp objects. The meter will switch itself off automatically after the test strip is ejected. Always use caution when removing the lancet.
METER MEMORY Your meter stores 450 the most recent blood glucose test results along with respective dates and times in meter memory. To enter the meter memory, start with the meter switched off. Reviewing Test Results 1. Press and release will appear on the display.
Reviewing Blood Glucose Day Average Results 1. Press and release When appears on the display, keep pressing for 3 seconds until the blinking “ ” appears. Release and then your 7-day average result measured in general mode will appear on the display. 2.
Page 33
2. Exit the meter memory Keep pressing and the meter will turn off after displaying the last test result. NOTE Any time you want to exit the memory, keep pressing for 5 seconds or leave it without any action for 3 minutes. The meter will switch off automatically.
DOWNLOADING RESULTS ONTO A COMPUTER You can use your meter with a USB cable and the Health Care Soft- ware System to view your test results on your personal computer. To learn more about the Health Care Software System or to obtain a USB cable separately, please contact local customer services or the place of purchase for assistance.
MAINTENANCE Battery Your meter comes with one 3V CR2032 lithium battery. Low Battery Signal The meter will display one of the messages below to alert you when the meter power is getting low. 1. The symbol appears along with display messages: The meter is functional and the result remains ac- curate, but it is time to change the battery.
Page 36
CR2032 CR2032 CR2032 BATTERY BATTERY BATTERY NOTE Replacing the battery does not affect the test results stored in the memory. As with all small batteries, these batteries should be kept away from small children. If swallowed, promptly seek medical as- sistance.
Caring for Your Meter To avoid the meter and test strips attracting dirt, dust or other con- taminants, please wash and dry your hands thoroughly before use. Cleaning 1. To clean the meter exterior, wipe it with a cloth moistened with tap water or a mild cleaning agent, then dry the device with a soft and dry cloth.
For further information, please refer to the test strip package insert. Important Control Solution Information Use only CLEVER CHOICE control solutions with your meter. Do not use the control solution beyond the expiration date or 3 months after first opening. Write the opening date on the control solution vial and discard the remaining solution after 3 months.
SYSTEM TROUBLESHOOTING If you follow the recommended action but the problem persists, or error messages other than the ones below appear, please call your local customer service. Do not attempt to repair by yourself and never try to disassemble the meter under any circumstances. Result Readings MESSAGE WHAT IT MEANS...
Error Messages WHAT DOES WHAT IT MESSAGE ACTION METER SAY? MEANS Appears when the batteries Replace the bat- Battery is dead. cannot provide teries immedi- Please replace. enough power for ately. a test. Test strip has Appears when a Repeat the test been used.
Troubleshooting 1. If the meter does not display a message after inserting a test strip: POSSIBLE CAUSE WHAT TO DO Battery exhausted. Replace the battery. Test strip inserted upside down or Insert the test strip with contact bars incompletely. end first and facing up. Defective meter or test strips.
DETAILED INFORMATION Reference Values Blood glucose monitoring plays an important role in diabetes control. A long-term study showed that maintaining blood glucose levels close to normal can reduce the risk of diabetes complications by up to 60%* . The results provided by this system can help you and your healthcare professional monitor and adjust your treatment plan to gain better control of your diabetes.
Comparing Meter and Laboratory Results The meter provides you with whole blood equivalent results. The result you obtain from your meter may differ somewhat from your laboratory result due to normal variation. Meter results can be af- fected by factors and conditions that do not affect laboratory results in the same way.
Page 44
You may still have a variation from the result because blood glucose levels can change significantly over short periods of time, espe- cially if you have recently eaten , exercised, taken medication, or experienced stress* . In addition, if you have eaten recently, the blood glucose level from a finger prick can be up to 70 mg/dL (3.9 mmol/L) higher than blood drawn from a vein (venous sample) used for a lab test*...
SPECIFICATIONS Model No.: TD-4267 Dimension & Weight: 90mm(L) x 55mm (W) x 18mm (H), 49g (without battery) Power Source: one CR2032 lithium battery Display: LCD Memory: 450 measurement results with respective date and time External output: mini USB cable Auto electrode inserting detection Auto sample loading detection Auto reaction time count-down Auto turn-off after 3 minutes without action...