Restoring The Factory Default Setups - Setra Systems L Series User Manual

Precision balance
Table of Contents

Advertisement

DISABLING TONE
Your Setra balance emits a tone when a number or function is entered. To disable the
tone, use the following procedure:
1.
Press the — key.
2.
Enter 9999. This is the tone access code.
3.
Press the cal key. The tone is disabled.
NOTE: To return the tone to use, repeat Steps 1-3.
FIRMWARE VERSION
Setra is continually making improvements to its products. As a result the operating
software, or firmware, of the scale has a revision number. The revision number may
prove useful in communicating with your dealer or Setra in case of difficulty.
To find out your scale's firmware version number, follow this procedure:
1.
Press the — key and enter 7777.
2.
Press the setup/cal key. The scale will display "rEF x.x", where "x.x" is firmware
version number.
3.
Press clear to return the scale to normal operation.

RESTORING THE FACTORY DEFAULT SETUPS

The many features described in this section allow the user to customize the scale to suit
a particular application. However, in doing this it is also possible to inadvertently setup
the scale in such a way that it will not operate as expected. To reset the scale to the
original factory defaults, follow the steps outlined below.
The factory defaults are set as follows:
-
Baud rate is equal to 2400 and no parity.
-
print is assigned to the standard serial tape printer (Setra 240P).
-
The beeper is enabled.
-
The int mode is set to rapid display update (short integration time).
-
Span = factory calibration
-
X units = 1 gram.
-
T units = 1 grain.
To restore the factory default setups use the following procedure:
1.
Press the — key and enter 8888.
18

Hide quick links:

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the L Series and is the answer not in the manual?

This manual is also suitable for:

12000l5000l2000l500l