GasAlertMicroClip XT
Technical Reference Guide
Sensor Alarm
The detector tests for missing or defective sensors during the startup
self-test and continuously thereafter.
• During start-up if a sensor fails, Error and the sensor name dis-
plays.
• If a sensor fails the self-test, Err displays above the gas type of the
failed sensor. The detector will enter alarm. Press C to acknowl-
edge the sensor alarm and stoop the alarm. Refer to
ing.
Low Battery Alarm
• Battery power is continually displayed during normal operation. If
battery power is low, L and
forms a sequence of 10 rapid sirens and alternating flashes with
7 seconds of silence in between (continues for 15 minutes).
36
Troubleshoot-
flashes. The detector per-
• If battery power becomes critically low, L and
play. The detector performs a sequence of 10 rapid sirens with 1
second of silence in between (sequence reactivates seven times).
The detector then displays
Charge the battery immediately. Refer to
Confidence Beep automatically disables during a low battery
alarm.
Automatic Deactivation Alarm
An automatic deactivation alarm will occur if
• the battery voltage is too low to operate the detector,
• calibration is due but not performed (when the Force Calibration
option is enabled),
• bump test is due but not performed (when the Force Bump
options is enabled) and
• all sensors fail during the startup self-test.
The detector performs a sequence of 10 rapid sirens with alternating
flashes with 1 second of silence in between (sequence reactivates
seven times). OFF then displays and the detector deactivates.
LOW BAT
OFF
and the detector deactivates.
Charging the
Battery.
Note
dis-