Calibration; Calibration Interval; Test Equipment Requirements - Pacific AMX series Operation Manual

Amx series ac power source
Table of Contents

Advertisement

6.0

CALIBRATION

This section describes the calibration of the AMX-Series AC Power Source.
6.1

CALIBRATION INTERVAL

The AMX-Series Power Source requires calibration once every six months or after service has been performed
to the system.
6.2

TEST EQUIPMENT REQUIREMENTS

The test equipment listed below is required for calibration of the AMX-Series Power Source.
1.
2.
3.
4.
5.
SECTION 6
CALIBRATION
Digital Voltmeter:
Frequency counter:
Digital Ammeter: 3½ Digit True-RMS responding
(Alternate Approach: Current transformer used in conjunction with the DVM.)
Oscilloscope:
Load:
4½ Digit True-RMS responding
5000 Hz bandwidth, min.
5 digit counter, min.
5000 Hz bandwidth, min.
(Optional)
Varies by model number;
2A @ 125 VAC, Model 305, 308-AMX
2.6A @ 150 VAC, Model 312-AMX
3A @ 125 VAC, Model 105, 108-AMX
4A @ 150 VAC, Model 112-AMX
6A @ 125 VAC, Model 320-AMX
10A @ 125 VAC, Model 125-AMX
12A @ 125 VAC, Model 345-AMX
16A @ 125 VAC, Model 140-AMX
16A @ 125 VAC, Model 160-AMX
16A @ 125 VAC, Model 360-AMX
24A @ 125 VAC, Model 390-AMX
32A @ 125 VAC, Model 3120-AMX
101

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents