Limitations

- Store the GE100 Blood Glucose Test Strips in the original capped vial at temperatures between 39
86 ( 4
°F
°C
to 30
°C
) and relative humidity below 90%. Do not freeze.
- GE100 Blood Glucose Test Strips are designed for using with capillary whole blood samples. Do not use
serum or plasma samples.
- Inaccurate test results may be obtained at altitudes greater than 10,000 feet (30048 meters) above sea
level.
- Hematocrit below 30% may cause higher results, and hematocrit above 60% may cause lower results.
- Severe dehydration and excessive water loss may cause inaccurately low results.
- GE100 Blood Glucose Monitoring System has not been validated for use on neonates.
- Inaccurate test results may be obtained by abnormal concentrations of
Uric acid ≧ 10.0 mg/dL, L-Dopa ≧ 3 mg/dL, Dopamine ≧ 2 mg/dL, Ascorbic acid (Vitamin C)≧ 5 mg/dL
NOTE
- Acetaminophen, Ibuprofen, Methyldopa, Salicylic Acid, Tetracycline, Tolbutamide, Bilirubin-
unconjugated, Cholesterol, Creatinine, Triglyceride, Maltose, Xylose, Galactose and Lactose
(when occurring in normal blood or normal therapeutic concentrations) do not significantly
affect results. However, abnormally high concentrations in blood may cause inaccurate results.
- Sodium Fluoride/Potassium Oxalate at low glucose levels may result in inaccurate blood glucose
reading. Sodium Fluoride/Potassium Oxalate should not be used as an anticoagulant with this
device.
46
°F
to
Measurement Technology
Measurement Calibration
Sample
Minimum Sample Volume
Measuring Range
Test Time
Memory Capacity
Power Saving
Operating Temperature
Operating Relative Humidity

Specifications

Oxidase Electrochemical Sensor
Plasma
Capillary whole blood
1.0 microliters
20-600 mg/dL
5 seconds
500 blood glucose test results with date and time
Turns off automatically 2 minutes after last user
action or press the "
" button for 3 seconds.
50 - 104
°F
(
10 - 40
°C
)
10 - 90%
47