3. When the U-TV0.5/C-2/U-TV0.5/C-3/U-TV0.63XC-1-2(C mount camera adapter) is used, using two or more
intermediate attachments* may obscure the peripheral part of the field of view or may make flare noticeable.
* Example of two intermediate attachments with BX microscope :
Reflected light illuminator + intermediate attachment equivalent to the U-CA.
4. Under fluorescent ring illumination or other AC-driven illumination such as a phase control light intensity
adjusting illumination system, the following phenomena may be observed by illumination light flickering
because of high light intensity and shortened exposure time :
· Flickering of the displayed image.
· Instability in exposure.
· Hatching patterns in pixel shift recording (4800 x 3600 or 2400 x 1800 pixels, 3CCD mode).
However, provided that the brightness can be adjusted using the light intensity control knob or ND filters, the
above phenomena may be attenuated by adjusting the brightness so that the exposure time exceeds 1/50
sec.
For details on the microscope models using AC-driven illumination, contact Olympus.
5. Combinations of this product and non-Olympus microscopes have not been evaluated extensively.
Non-Olympus microscopes and commercially available C-mount lenses can be used provided that they
match a CCD with a size of no less than 1/1.8 inch and the lens projection length from the C-mount body
attaching section is no more than 4.5 mm. However, problems due to optical adaptability, such as shading,
may be observed.
DP73