Assembly Instructions; Before Assembly; Assembling The 2210Kl; Operation Instructions - Health O Meter 2210KL User Instructions

Neonatal pediatic scale, 1 gram accuracy
Hide thumbs Also See for 2210KL:
Table of Contents

Advertisement

Calibration
Your scale has been properly calibrated at the factory and should not require any further
adjustments. The following procedure should be performed only to correct any error in
calibration. Calibration is protected by a security code to avoid inadvertent access. For
maximum accuracy, calibration must be done in grams. If you only have certified weights in
pounds, convert them to grams (i.e. 10 pounds = 4536 grams).
1. Depress the ON/ZERO button on the console and wait for the display to show 0.0 .
2. Place an accurately known sure weight (3,000 to 5,000 grams. recommended) on the
platform. If the display reads other than the known weight value by +/- 1 gram, then the
scale is out of calibration.
IMPORTANT NOTE: A sure weight is a weight that has been accurately calibrated and its license
is traceable to NBS weights & measures. Weight room free weights are not normally suitable to use
as check weights as they may vary from their stated value. If you do not have access to sure
weights , use a weight that has been weighed on a scale you are sure is accurate.
3. Remove weight from platform. The calibration security code must be entered before
attempting to calibrate the scale. Hold the MENU button down and press RECALL button.
Release both buttons. Set-0 will be displayed and the 0 will be flashing.
button repeatedly until Set-2 appears. Depress the ENTER button.
4. Depress the
keys to input 125 as the calibration security code. Press ENTER to
5. Use the
and
advance to the next digit. After entering the third digit, press ENTER and the scale will
return to normal weighing mode. The security code needs to be entered only once, unless
the scale is powered off. Access to calibration is denied if the proper code is not entered.
6. With the known weight on the scale, press REWEIGH. Quickly, before WeightLock
activates, hold the MENU button down and press RECALL button. Release both buttons.
Set-0 will be displayed and the 0 will be flashing.
button repeatedly until Set-6 appears. Depress the ENTER button.
7. Depress the
8. A weight value will display representing the current calculation of the known weight. The
hundreds of pounds digit will be flashing. Use the
value. Press ENTER to advance to the next digit. After entering the tenths of pounds digit,
press ENTER and -CAL- will display while the scale calibrates itself.
9. The readout will then quickly show the old weight, then re-weigh to show the new
calibration weight.
10.Remove weight and depress the ON/ZERO button. Wait for 0.0 to display and verify
calibration by placing the same known weight back on the platform.
Calibration Error Code
Err-4
Scale is below minimum calibration weight. Use at least 1000 grams to
calibrate the scale.
and
keys to input the proper
12

ASSEMBLY INSTRUCTIONS

Before Assembly

Carefully inspect the carton for shipping
damage before unpacking. If significant
damage is found, keep all packaging and
contact your shipper or Pelstar immediately.
Claims must be filed with the shipper as soon
as possible after receipt of damaged goods.
Carefully unpack the components from
the shipping containers. Your scale has
been shipped in one (1) carton and contains
the following components:
·
Scale Platform
·
Acrylic Cradle
Six C-cell alkaline batteries are pre-installed
in the scale platform
Length rule
Figure 1: Install Cradle

Assembling the 2210KL

1. Place the scale platform on a table top
with the display buttons facing you.
2. The acrylic cradle has a built-in length
rule. Align the cradle so that the rule is
positioned over the back end of the scale
platform (Fig. 1).
3. Carefully lower the cradle onto the four
mounting pins on the scale platform.
When the cradle is properly aligned with
the pins, gently push down on the cradle
over each pin until it snaps into its locked
position. When properly positioned, there
will be about a 3/8 space between the
platform top and the bottom of the cradle.
Always lift the scale by the black handles
only
never by the cradle.

OPERATION INSTRUCTIONS

Quick Start Instructions

1. If desired, place blanket on cradle
2. Press blue ON/ZERO button. Make sure
0 STABLE appears on display.
3. Place baby on cradle.
4. Allow scale to lock onto the weight. Prior
Weight will show on the display.
5. Remove baby from cradle. Measurement
value remains on the display.
6. Press blue ON/ZERO button to ready
scale for next weighing.
5