Using the Tare Function

When using this scale, the weight of any
object, such as a blanket, accompanying the
baby being weighed can be subtracted from
the total weight to determine the net weight
of the baby alone. The Tare function
automatically performs this subtraction,
avoiding the need for manual calculations.
TARE
Figure 3: Tare button
gram
Figure 4: Net weight display
1. Turn the scale on and wait for the display
to read 0 .
2. Place the object (i.e. blanket) on the
cradle. The weight of the object will be
displayed.
3. Depress the TARE button. NET WT and
0.0 will be displayed .
If the object is removed from the platform,
its weight will be displayed as a negative
value.
4. Place the baby on the cradle with the
object still on it. The readout will now
display the baby s net weight.
5. Depress ON/ZERO to return the scale
back to normal weighing operation.
Using the Weightlo ck Function
When using this 2210KL scale, the weight of
any object over 512 grams will be locked on
the display even after the baby has been
removed from the cradle allowing the care
provider to tend to the needs of the baby
without having to remember the weight
measurement. This value is stored and can
be recalled even after the scale has turned
itself off.
When the baby s weight is locked, PRIOR
WT is displayed (Fig. 5). The weight will
remain displayed even after the baby is
removed from the cradle. No further
weighings can take place until Weightlock is
deactivated.
Weightlock is deactivated by depressing
ON/ZERO or REWEIGH/RECALL (Fig 6).
gram
Figure 5: Weightlock display
REWEIGH
RECALL
Figure 6: Reweigh/Recall button
8
Using the Reweigh Function
The Reweigh function is used to reweigh the
baby while he or she is still on the cradle.
To reweigh, the baby should remain on the
cradle and a weight measurement should be
on the display. Simply press REWEIGH/
RECALL.
Using the Recall Function
The Recall function is used to recall the last
weight measurement after the scale has shut
down and the display is blank.
The scale automatically stores the last weight
measurement. This value is stored until a
new measurement is taken or it is
intentionally cleared.
To recall a weight value:
1. The baby s weight has been measured,
the baby has been removed from the
cradle and the display is blank (scale has
shut down).
2. Depress ON/ZERO and then REWEIGH/
RECALL. The stored value will now
display.
3. With the weight measurement in the
display, the scale will be returned to the
unlocked mode (ready for weighing) by
either of the following:
a. RECALL is depressed again
b. ON/ZERO is depressed
Programming Auto-Off Time
The 2210KL has been preset at the factory to
shut down after one minute of no use. This
can be programmed from 12 seconds to 25
minutes to suit your needs.
1. Depress ON/ZERO and wait for the
display to show 0.0 . Do not place
anything on cradle.
2. Hold MENU button down and depress the
RECALL button. Set-0 will be displayed
along with the up and down arrows; the
0 will be flashing.
3. Depress the
button repeatedly until
Set-3 appears, depress ENTER.
4. The Auto-Off time currently programmed
in the scale will display as minutes and
tenths of minutes (i.e. 03.2 ); the far left
digit will be flashing. Use the
and
buttons to set the desired value, press
ENTER to advance to the next digit. For
example, setting the time to 00.5 will turn
the scale off after 30 seconds of no use;
02.0 will turn the scale off after 2
minutes; 25.0 will turn the scale off after
25 minutes.
5. Press ENTER when complete to save and
return to normal weighing operation.
ENTER
MENU
Figure 7: Menu button