Keithley 7706 User Manual page 41

Multiplexer card
Table of Contents

Advertisement

Table 8
Model 7706 calibration commands
Command
:CALibration
:PROTected
:CODE '<up to 8 char. string>'
:CARD1
:INITiate
:COUNt?
:RCOunt
:DATE?
:SAVE
:LOCK
:LOCK?
:STEP0 <NRf>
:DAC
:STEP0 <NRf>
:CARD2
:COUNt?
:DATE?
:DIAGnostic
:KEIThley
:CALibration
:UNLOCK
Description
Calibration root command.
All commands in this subsystem are protected by the calibration lock (except
queries and :CODE).
Send calibration code. (Default KI002700)
Path to card in slot 1.
Initiate card calibration.
Request number of times card had been calibrated.
Reset card calibration count to 0.
Request card calibration date.
Save cal constants to card EEPROM.
Lock out calibration.
Request cal lock state. (0 = locked, 1 = unlocked)
Temperature sensor cold calibration (<NRf> = temperature, °C).
Path to analog output calibration.
Analog output DAC calibration.
Path to card in slot 2.
Request number of times card has been calibrated.
Request card calibration date.
Diagnostic root.
Keithley path.
Calibration path.
Unlock command.
Calibration commands
37

Advertisement

Table of Contents
loading

Table of Contents