14. Close channel 2:
a. Press the CLOSE key.
b. Key in 102.
15. Set analog output 2 to +10V:
a. Open the SHIFT-CARD menu (press SHIFT-CARD).
b. Key in analog output for AOUT2:+10.000V.
16. Verify analog output reading is within stated limits.
17. Set analog output 2 to -10V:
a. Open the SHIFT-CARD menu (press SHIFT-CARD).
b. Key in analog output for AOUT2:-10.000V.
18. Verify analog output reading is within stated limits.
19. Open channel 2 (press the OPEN key).
Figure 24
Verification—analog connections
Table 6
Analog output limits
Source DC Voltage
+10V
-10V
Reading limits (1 year, 18° to 28°C)
+9.966 to +10.034V 0.15% of reading + 19mV
-10.034 to -9.966V
Verification
31