Specifications; Connection Log - Keithley 7706 User Manual

Multiplexer card
Table of Contents

Advertisement

Model 7706 module
Table 1
Monitor scan example (front panel and remote steps)
Step Front panel operation
1
Restore defaults:
Restore defaults (SHIFT SETUP > RESTORE: FACT).
2
For front panel operation, proceed to step 3.
For remote programming, clear buffer and disable buffer auto clear:
3
Configure advanced scan (SHIFT CONFIG > ADVANCED):
a
Channel 101, 102, and 103:
Select DCV function.
Select 10V range.
Set filter count to 20 (SHIFT TYPE > 020 RDGS).
Enable filter (FILTER).
b
Channel 104:
Disable (off) Channels 104-120 (SHIFT CH-OFF).
c
Channel 125:
Enable Totalizer channel (SHIFT-CH OFF).
Set and enable high limit 1:
Set limit to 100000 (SHIFT LIMITS > HI1:+100.0000K).
Set and enable high limit 2:
Set limit to 1000000000 (SHIFT LIMITS > HI2:+1000000K).
Enable (on) limit (SHIFT OFF/ON > LIMITS: ON).
d
Disable immediate scan (IMM SCAN: N), and enable high limit 1
(HLIM1 SCAN:Y).
e
Disable timer (TIMER? OFF).
f
Set reading count to 4.
4
For front panel operation, proceed to step 5.
For remote programming, set the number of points in the monitor scan.
5
Select and enable monitor channel (SHIFT MONITOR > 125).
Specifications
Full Model 7706 specifications are included at the end of this guide.

Connection log

Make a copy of
Table 2
and affix it to the cover of the Model 7706. Use this to record connection information and channel de-
scriptions as needed.
24
Remote programming
*RST
TRAC:CLE
FUNC 'VOLT', (@101:103)
VOLT:RANG 10,(@101:103)
VOLT:AVER:COUN 20,(@101:103)
VOLT:AVER:STAT ON,(@101:103)
ROUT:SCAN (@101:103,125)
CALC3:LIM1:UPP 1e5,(@125)
CALC3:LIM1:STAT ON,(@125)
ROUT:SCAN:TSO HLIM1
ROUT:MON:POIN 4
ROUT:MON (@125)
ROUT:MON:STAT ON