Lakeshore 340 User Manual page 135

Hide thumbs Also See for 340:
Table of Contents

Advertisement

MDAT?
Query Min/Max Data for an Input
Input:
MDAT? <input>
Returned: <min value>,<max value>. Format: ±nnn.nnnE±n,±nnn.nnnE±n[term]
Remarks: Returns the minimum and maximum input data. <input> specifies which input to query.
MDATST?
Query Min/Max Data Status for an Input
Input:
MDATST? <input>
Returned: <min bit weighting>, <max bit weighting>. Format: nnn,nnn[term]
Remarks: The integer returned represents the sum of the bit weighting of the input minimum /maximum
data status flag bits. <input> specifies which input to query. See RDGST? Command.
MNMX
Configure Minimum and Maximum Input Function Parameters
Input:
MNMX <input>, [<on/pause>], [<source>]
Returned: Nothing
Remarks: Configures the minimum and maximum input functions.
<input>
<on/pause> Turns on or Pauses the max/min function. Valid entries: 1 = on, 2 = paused.
<source>
Example: MNMX B, 1, 3[term] - Input B min/max function is on and processes data from the input
sensor units reading.
MNMX?
Query Minimum and Maximum Input Function Parameters
Input:
MNMX? <input>
Returned: <on/pause>, <source>. Format: n,n[term]
Remarks: Returns a min/max input configuration. See the MNMX command for parameter descriptions.
<input> specifies which input to query.
MNMXRST
Resets Min/Max Function for All Inputs
Input:
MNMXRST
Returned: Nothing
Remarks: Resets the minimum and maximum data for all inputs.
MODE
Configure Remote Interface Mode
Input:
MODE <mode>
Returned: Nothing
Remarks: Configures the remote interface mode. <mode> specifies which mode to operate. Valid
entries: 1 = local, 2 = remote, 3 = remote with local lockout.
Example: MODE 2[term] - Places the Model 340 into remote mode.
MODE?
Query Remote Interface Mode
Input:
MODE?
Returned: <mode>. Format: n[term]
Remarks: Returns the remote interface mode. See the MODE command for parameter descriptions.
MOUT
Configure Control Loop Manual Output Value
Input:
MOUT <loop>, <value>
Returned: Nothing
Remarks: Configures the control loop manual output values. <loop> specifies loop to configure.
<value> specifies value for manual output.
Example: MOUT 1, 22.45[term] - Control Loop 1 manual output is 22.45%.
Remote Operation
Lake Shore Model 340 Temperature Controller User's Manual
Specifies input to configure.
Specifies input data to process through max/min. Valid entries: 1 = kelvin,
2 = Celsius, 3 = sensor units, 4 = linear data.
9-37