Table 12. Electromagnetic compatibility tests
Description
1 MHz burst disturbance test:
• Common mode
• Differential mode
Electrostatic discharge test:
• Contact discharge
• Air discharge
Radio frequency, electro-magnetic
field immunity test:
Fast transient disturbance tests:
• All ports
Surge immunity test:
• Common mode
• Differential mode
Immunity to conducted disturbance
induced by RF:
• Continuous
• Short duration ( 1 sec )
GUID-89A9F7DF-FB82-4E8D-B449-7C0BB51C1DCA V1 EN
Conducted radio frequency
interferene tests:
Pulse magnetic field immunity tests:
10
Type test value
2.5 kV, 1MHz, 400 pulses/sec
1.0 kV, 1MHz, 400 pulses/sec
Ω
6 kV, 150 pF/330
8 kV, 150 pF/330
Ω
10 V/m
f=80-1000 MHz
10 V/m
f=80, 160, 450, 900 MHz
4 kV, 5.0 kHz
1.0 kV, 1.2/50 μs
0.5 kV, 1.2/50 μs
100 A/m
1000 A/m
10 V
f=150 KHz...80 Mhz
1000 A/m, 6.4/16 μs
1MDB07207-YN
Reference
IEC 61000-4-12
IEC 60255-22-1, class III
IEC 60255-22-2, class III
IEC 61000-4-2
IEC 60255-22-3, class III
IEC 61000-4-3
IEC 60255-22-4, class A
IEC 61000-4-4
IEC 60255-22-5
IEC 61000-4-5
IEC 61000-4-8
IEC 60255-22-6, class III
IEC 61000-4-6
IEC 61000-4-9