Surgical instruments (OP)

Surgical instruments should be stored
for as short a time as possible before
machine processing.
The OXIVARIO programme is
recommended for instruments where
there is a long delay between the time
they are used and the time they can be
reprocessed. See "Special features" at
the end of this booklet.
Disinfection of surgical instruments,
and of those used for minimal invasive
surgery should preferably take place
thermally.
De-ionised water should be used for
the final rinse whenever possible to
ensure no marks are left on the load
and to avoid corrosion. If the water
used contains more than 100 mg
chloride/litre there is danger of
corrosion.
OP-containers can be disinfected
thermally in the CONTAINER
programme, or chemo-thermally. If
anodised aluminium containers are
used the final rinse must be carried out
with de-ionised water. These containers
must not be processed with an alkaline
cleaning agent in a 93 °C programme
with 10 minutes holding time
(according to § 18 IfSG).
Mobile units for surgical instruments
and containers are supplied with
their own operating instructions.
Areas of application
When cleaning narrow lumen
instruments e.g. those used for
minimally invasive surgery, an
intensive internal cleaning result is
imperative. The Vario-TD and
OXIVARIO programmes are the only
ones which offer a thorough enough
level of cleaning for this. It is essential
that instruments are loaded as directed
and that the cleaning agent used is
suitable for the programme and for the
sensitive instruments being processed.
The final rinse must be carried out
using fully demineralised water with a
conductivity level of ~15 ZS/cm
(microsiemens per centimeter).
23