Invacare Top End Terminator Jr. Operating And Maintenance Manual page 21

Invacare top end owner's operator and maintenance manual
Hide thumbs Also See for Top End Terminator Jr.:
Table of Contents

Advertisement

ARMS
2. Slide the T-arm to one (1) of:
A. Low Height T-Arms - Nine (9) positions.
B. High Height T-Arms - Seven (7) positions.
NOTE: If the inside T-arm post does not slide up and
down in the outside T-arm post as desired, perform
one (1) of the following:
A. Tighten - Tightening the set screws on the out-
side T-arm post will make it harder to move the
inside T-arm post up and down.
B. Loosen - Loosening the set screws on the out-
side T-arm post will make it easier to move the
inside T-arm post up and down.
3. Lock the T-arm by flipping the T-arm release le-
ver towards the front of the wheelchair.
Set
Screws
Outside
T- Arm Post
T-Arm Release Lever - Locked Position
FIGURE 8 - ADJUSTING T-ARM HEIGHT
Adjusting T-Arm Width (FIGURE 9)
1. Remove the two (2) phillips screws that secure the
arm pad to the arm tube.
2. Turn the arm pad around and reposition the arm pad
on the arm tube.
3. Re-secure the arm pad to the arm tube with the two
(2) phillips screws. Tighten securely.
4. Repeat for the opposite side, if necessary.
Arm Pad
Phillips
Screw
FIGURE 9 - ADJUSTING T-ARM WIDTH
Inside T-Arm
Post
T-Arm Release Lever -
Unlocked Position
Arm Tube
Phillips Screw
Adjusting T-Arm Depth (FIGURE 10)
1. Remove the two (2) phillips screws that secure the
arm pad to the arm tube.
2. Remove the two (2) socket screws that secure the
arm tube to the T-arm post.
3. Reposition the arm tube on the T-arm post:
A. Desk Length Arms - to one (1) of three (3) posi-
tions depending on the desired arm pad depth.
B. Full Length Arms - to one (1) of five (5) posi-
tions depending on the desired arm pad depth.
NOTE: Additional positions are obtainable by turning
o
the arm tube 180
.
4. Re-secure the arm tube to the T-arm post with the
two (2) socket screws. Torque to 60-70 in./lbs.
5. Reattach the arm pad to the arm tube with the two (2)
phillips screws. Tighten securely.
6. Repeat for the opposite side, if necessary.
Socket
Screws
Phillips
Screw
T-Arm Post
FIGURE 10 - ADJUSTING T-ARM DEPTH
Adjusting T-Arm Sockets (FIGURE 11)
NOTE: Perform this procedure if the T-arm is too loose
in the socket or does not easily slide up and down in
the socket.
1. Remove the rear wheels from the wheelchair, if nec-
essary. Refer to REMOVING/INSTALLING REAR
WHEELS in PROCEDURE 7 of this manual.
2. Loosen, but do not remove the four (4) hex screws
and washers that secure T-arm socket to T- arm clamp.
NOTE: The T-arm socket will disassemble if the four
(4) hex screws and washers are removed.
21
PROCEDURE 5
A
R
M
S
Arm Pad
Arm Tube
NOTE: If
necessary, turn
o
arm tube 180
to obtain two
(2) positions.
Phillips Screw

Advertisement

Table of Contents
loading

This manual is also suitable for:

Top end terminator junior