Section 4-Typical Product Parameters - Invacare IRC10LX Operator's Manual

Platinum series oxygen concentrators standard, with senso2 and homefill
Hide thumbs Also See for IRC10LX:
Table of Contents

Advertisement

SECTION 4—TYPICAL PRODUCT
PARAMETERS
Electrical Requirements:
Rated Current Input:
Sound Level:
Altitude:
*Oxygen Output
Concentration Levels:
Maximum Outlet Pressure:
*NOTE: Stated concentration levels achieved after initial warm‐up 
period (approximately 30 minutes).
Part No 1118353
SECTION 4—TYPICAL PRODUCT PARAMETERS
120 VAC ± 10% (132 VAC/108 VAC), 60
Hz
Platinum 5 and XL: 4.3 A
Platinum 10: 4.7 A
Platinum 5 and XL: 50 dB Average
Platinum 10: 58 dB Average
Platinum 5 and XL - Up to 6,000 ft (1828
meters) above sea level without degradation
of concentration levels. From 6,000 ft (1828
meters) to 13,129 ft (4000 meters) below
90% efficiency
Platinum 10 - Up to 4,000 ft (1230 meters)
above sea level without degradation of con-
centration levels. From 4,000 ft (1230
meters) to 13,129 ft (4000 meters) below
90% efficiency
All 5LXO2/5LX models 87% to 95.6% for
flows ranging 0.5 to 5 L/min and all
10LXO2/10LX models 87% to 95.6% for
flows ranging 2 L to 10 L/min
Platinum 5 and XL: 5 psi ± 0.5 psi (34.5 kPa
± 3.45 kPa)
Platinum 10: 9 psi ± 0.5 psi (62.0 kPa ± 3.45
kPa)