Technical Data; Typical Product Parameters - Invacare Perfecto2 Series User Manual

Homefill system compatible oxygen concentrators
Hide thumbs Also See for Perfecto2 Series:
Table of Contents

Advertisement

Available languages

Available languages

Invacare® Perfecto₂™Series

10 Technical data

10.1 Typical Product Parameters

Electrical Requirements:
Rated Current Input:
Sound Level:
Altitude:
Oxygen Output Concentration Levels:
Maximum Outlet Pressure:
Low Flow Alarm:
Potential Obstruction Alert:
30
120 VAC + 10, –15% (132 VAC/102 VAC), 60 Hz
3 A
43 ± 2 dBA Average for Perfecto
40 ± 2 dBA Average for Perfecto
All Perfecto
Models:Up to 8,000 ft (2438 m) above sea level
2
without degradation of concentration levels.
All Perfecto
Models while filling a HomeFill System:Up to
2
6,000 ft (1828 m) above sea level without degradation of concentration
levels.
Atmospheric Pressure Range: 101.3 kPa – 75.0 kPa
87% to 95.6% at 0.5 to 5 L/min
Concentration levels achieved after initial warm-up period
(approximately 30 minutes)
5 psi ± 0.5 psi (34.5 kPa ± 3.45 kPa)
0.5 to 5 L/min (maximum). For flowrates less than 1 L/min, we
recommend the use of the Invacare Pediatric Flowmeter Accessory
(IRCPF16).
0 L/min to 0.5 L/min
The concentrator detects a condition that may indicate a potential
obstruction of the output oxygen. Rapid audible beeping alert (this alert
is deactivated when accessories are connected). May be associated
with flow setting of 0.5 L/min or less.
Models and Perfecto
V Model
2
2
W Models
2
1148074-F-06

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Irc5po2Irc5po2vIrc5pIrc5pwIrc5po2w

Table of Contents