Flowrate - Invacare IRC5PO2V User Manual

Perfecto2 series oxygen concentrators
Hide thumbs Also See for IRC5PO2V:
Table of Contents

Advertisement

Flowrate

NOTE: For this procedure, refer to FIGURE 5.5 on page 20.
1. Turn the flowrate knob to the setting prescribed by your 
physician or therapist.
Never set the flow greater than 5 L/min.
DO NOT change the L/min setting on the flowmeter unless a 
change has been prescribed by your physician or therapist.
NOTE: To properly read the flowmeter, locate the prescribed flowrate 
line on the flowmeter. Next, turn the flow knob until the ball rises to 
the line. Now, center the ball on the L/min line prescribed. 
2. If the flowrate on the flowmeter ever falls below 0.5 L/min 
for more than about one minute, the LOW FLOW alarm 
will be triggered. This is a rapid beeping of the audible 
alarm. Check your tubing or accessories for blocked or 
kinked tubing or a defective humidifier bottle. After rated 
flow is restored, between 0.5 L/min and 0.75 L/min, the 
LOW FLOW audible alarm will turn off.
NOTE: The use of some accessories such as the pediatric flowstand 
and the HomeFill compressor will deactivate the Low Flow Alarm.
Part No 1143482
SECTION 5—OPERATING INSTRUCTIONS
CAUTION
WARNING
19
Perfecto
Series
2

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Irc5po2wIrc5pIrc5po2Irc5pw

Table of Contents