Recommended Test Equipment - Agilent Technologies U8001A User's And Service Manual

Single output dc power supplies
Hide thumbs Also See for U8001A:
Table of Contents

Advertisement

5
Service Guide

Recommended Test Equipment

Table 5-2 List of Equipment
Equipment
Digital Multimeter
(DMM)
Oscilloscope
Differential
Amplifier
RMS Volt Meter
50 Ω Feed-through
Termination
AC/DC Current
Converter
Electronic Load
AC Power Source
44
The test equipment recommended for the calibration and
adjustment is listed below. If the exact instrument is not
available, substitute the calibration standards of equivalent
requirement(s).
Requirement(s)
8½ digits resolution
• Sensitivity: 1 mV
• Bandwidth limit: 20 MHz
• Probe:
1:1 with RF tip
10:1 with RF tip for > 50 V
• Bandwidth > 20 MHz
(bandwidth limit : 20 MHz)
• AC coupling
• Amplifier x10
20 Hz to 20 MHz
• Voltage range: 30 Vdc
• Current range: 5 A
• Open and Short switches
• Transient On/Off
Capable of supplying
90 Vac to 250 Vac
Recommended Model Purpose
Agilent 3458A
Measures accurate
DC voltage
Agilent DSO8064A or
Displays transient
Infiniium equivalent
response. Displays
ripple and noise
waveform.
LeCroy DA1855A
Signal conditioning
preamplifiers
R&S URE3
Measures RMS noise
Pomona
Loads termination to
match output
impedance of the
differential amplifier
Tektronix TCP305 &
Converts current to
Tektronix TCPA300
voltage
Agilent 6060B
Measures load and
line regulations and
transient response
time
Agilent 6813B
Functions as variable
voltage transformer
U8001A/U8002A User's and Service Guide
Used
*
PV
C
PV
PV
PV
PV
PV
PV