N O T E
U8001A/U8002A User's and Service Guide
To disable the OCP
1 Press on the "Over Current" button to disable the OCP. The
OCP annunciator will eventually turn off.
To clear the over current condition
1 OCP trip can only occur when the output is enabled. When
the OCP condition occurs:
• The output is disabled and the OFF annunciator will be
turned on.
• If the keylock feature was enabled, it will be disabled. The
LOCK annuciator will be turned off.
• The OCP annunciator will fast- blink continuously.
• The display will show "TRIP".
2 To clear the OCP trip conditon, press on the "Over Current"
button again.
If OCP trip persists, decrease the current limit settings to clear the tripping.
Operations and Features
3
27