3
Operations and Features
Memory Operations
N O T E
22
For Agilent U8001A/U8002A single output DC power supplies,
up to three operating states can be stored in non- volatile
storage locations. The storage feature remembers the voltage
and current limit value settings, OVP
OVP/OCP trip levels. The following steps show you how to store
and recall an operating state.
Storing Operating State
1 Press on the "Memory" button.
• The M1 annunciator will blink.
2 Turn the knob to choose the memory location for storing
the operation state.
• The M1, M2 and M3 annunciators will blink in a
round- robin manner.
3 Press on the "Memory" button again to save the operating
state in the chosen memory.
• Display will show "donE".
• The M1, M2 and M3 annunciators will be turned off.
To cancel the operation, allow the unit to idle for five seconds.
* OVP- over voltage protection
† OCP- over current protection
*
†
/OCP
On/Off states, and
U8001A/U8002A User's and Service Guide