SECTION 7—TROUBLESHOOTING GUIDE
SECTION 7—TROUBLESHOOTING
GUIDE
SYMPTOM:
Alarm:
SHORT BEEPS,
LONG PAUSE
Concentrator
NOT operating,
power switch
ON.
Beep......Beep......
Alarm:
CONTINUOUS
Concentrator
NOT operating,
Power Switch
ON.
Beeeeeeeep....
Platinum
Series
PROBABLE CAUSE:
Main Power Loss:
1. Power cord NOT
plugged in.
2. No power at outlet.
3. Tripped circuit
breaker.
System Failure:
1. Unit overheating due
to blocked air intake.
2. Insufficient power at
outlet.
3. Internal repairs
required
1. Insert plug into outlet.
2. Inspect house circuit
breakers or fuses. If problem
recurs, use a different outlet.
3. Call Service Provider.
1a. Remove and clean cabinet
filters.
1b. Move oxygen concentra-
tor at least three (3) inches
away from walls, draperies or
furniture.
2. Do not use extension
cords. Move to another elec-
trical outlet or circuit.
3. Call Service Provider.
32
SOLUTION:
Part No 1118353