SECTION 5—OPERATING INSTRUCTIONS
Recommended Guidelines for Optimum
Performance
Temperature:
Electrical:
Placement:
Tubing and Cannula:
Environment:
Relative Humidity:
Time of Operation:
Flow Rate
Minimum Operating Time:
NOTE: Ensure that your concentrator is at least three (3) inches
away from walls, draperies or furniture to assure sufficient air flow.
Avoid deep pile carpets and heaters, radiators or hot air registers.
Plug in Power Cord
1. Plug in power cord to an electrical outlet.
Connect Humidifier (If So Prescribed)
NOTE: For this procedure, refer to FIGURE 5.1 - FIGURE 5.4.
Platinum
Series
™
50°F - 95°F (10°C - 35°C)
No extension cords.
No closer than three (3) inches from the
wall, furniture, draperies, or similar surfaces.
7 ft cannula with a maximum 50 ft of
Crush-Proof Tubing (DO NOT pinch).
Platinum 10: Recommended use up to 25 ft
high flow tubing with high flow cannula at all
flow rates.
Smoke and soot-free. No confined spaces
(Example: No closets).
20 to 60%
Up to 24 hours per day.
Platinum 5 and XL: From 0.5 L/min. to 5
L/min. For flow rates less than 1 L/min., use
the Pediatric Flowmeter Accessory.
Platinum 10: From 2 L/min. to 10 L/min
30 Minutes
22
Part No 1118353