Visual, Audio
Indicators
1 short beep
3 long beeps
+
1 short beep
Alternating short
and long beeps
Description
Low Battery Alarm
This alarm occurs
when approximately
17 minutes of
battery life remains�
Remaining battery life
is dependent on your
device settings and your
activity level� The empty
battery symbol flashes
on-screen�
No Flow Alarm
This alarm occurs when
the device detects there
is no oxygen flowing in
the patient cannula�
Depleted Battery
Alarm
This alarm occurs when
approximately two
minutes of battery
life remains� The
battery symbol flashes
on-screen�
Alarm Silence Symbol
This symbol appears
when you press the
Alarm Silence button
to silence the audible
alarm (beeps)�
SimplyGo User Manual
What To Do
Replace the battery
or connect to a power
source�
(Press the Alarm
symbol to silence the
alarm�)
Check the cannula
for kinks or other
obstructions that are
stopping oxygen flow
through it�
Replace or recharge
battery�
(Press Alarm symbol
to silence the alarm�)
No further action is
needed�