Performance Tests

To avoid electric shock, do not perform the performance test procedures unless
the Meter is fully assembled.
The following performance tests verify the complete operation of the Meter and check the accuracy of
each meter function against the Meter's specifications. If the Meter fails any part of the test, calibration
adjustment and/or repair is indicated.
In the performance tests, the Meter is referred to as the unit under test (UUT).

Required Equipment

A Fluke 5500A Multi-Product Calibrator (or equivalent) is required for the performance test procedures
in this document.
If an equivalent calibrator is used, it must meet the accuracy specifications shown in Table 1.
Recommended Equipment
5500A Multi-Product Calibrator
(or equivalent)
Fluke 80 AK
Thermocouple Adapter Accessory
K-type Thermocouple
10
WWarning
Table 1. Calibrator Specifications
Measurement Function
DC Volts
DC Current
AC Volts
AC Current
Resistance
Capacitance
Temperature
Frequency
K-type
K-type, mini-plug on both ends
Accuracy
0 to 1000 V
± 0.075 %
0 to 10 A
± 0.425 %
0 to 1000 V
± 0.375 % @ 45 Hz to 1 kHz
0 to 10 A
± 0.55 %
0 to 4 MΩ
± 0.125 %
4 to 40 MΩ
± 0.5 %
500 pF to 5 µF
± 0.325 %
.25%
0.5 Hz to 100 kHz
± 0.005 %
-
-