PROBLEM
CAUSE
Purified water does
Crimp in supply hose
not fill
Motor wires
disconnected
Tank and inner
Collects normally
door have surface
rust
Glassware is
contaminated with
mineral acids.
Washer does not
No power to washer
run
Door latch is open
Washer door opens
Cabinet opening is not
and closes badly
square
Door hitting counter-
top mounting screws
Washer leaks
Washer is not level
Door not sealing
Tank gasket loose
Loose connection at
hose clamps
Loose connection at
fill valve
Improper detergent
Obstruction in drain
line
Washer overfilling
Washer leaks
Improper installation
around door seal
Unit not level
Door opened during
operation
Product Service 1-800-522-7658
Chapter 7: Troubleshooting
CORRECTIVE ACTION
Refer to installation instructions for proper
purified water piping installation.
Reconnect wires.
Clean with stainless steel cleaner and soft cloth.
Pre-rinse glassware. Clean with stainless steel
cleaner and soft cloth.
Check fuse or circuit breaker, or door switch.
Latch washer door by pushing door in.
Correct opening with leveling feet.
Adjust mounting screws.
Adjust leveling feet.
Adjust door latch.
Check gasket and correct.
Tighten hose clamp.
Check fill valve and correct.
Use non-foaming detergent. Use only proper
amount of detergent.
Remove obstruction.
Check operation of fill switch.
Check cabinet opening to see if it is aligned
properly with the door. Washer must be level.
Adjust feet to level unit.
Adjust feet to level unit.
Allow 4-5 seconds before opening door after
unlatching.
47