The 16-place Pipet Insert, shown below accommodates 1 ml to 50 ml pipets. Two
of these inserts may be installed, if desired. Refer to the figure and follow the
instructions below to install a 16-place Pipet Insert.
To install a 16-place Pipet Insert:
1. Use the tool supplied with the Spindle Rack or a screwdriver to remove the six
spindles on either the right or left side of the spindle rack. Remove others if
necessary.
2. Replace the six spindles with the six short spindles that you received with the
Pipet Insert.
3. Replace all other spindles with the plastic plugs that you received with your
washer.
4. With the support facing the center of the spindle rack, slide the Pipet Insert
over the spindles until it completely seats.
5. Push pipets into the 16 receptacles and rest them on the support.
Note: Lower spindle racks are shipped with 34 long spindles and 2 shorter
spindles. If spindles are removed and then replaced, the shorter spindles should
be located on the front row in the second and third positions from the right side.
This will allow proper function of the detergent dispenser.