Optex SS Checker SC-T4 Instruction Manual page 11

Hide thumbs Also See for SS Checker SC-T4:
Table of Contents

Advertisement

3
Signal Output Range Setting
Signal Output Range can be set as follows.
"High Limit" "Low Limit" 200
Setting unit is 5 mg/l.
The Signal Output Range is default at 4mA for the 0-Adjustment
value and 20mA for the Span-Adjustment value.
Ex: To set the The Signal Output Range for 100
( 1 ) Signal Output Low Limit Setting
1
Press
and select the
Output Low
Limit.
"-Lo-" and "the setting value" comes on
alternatively.
2
Press
for 2 seconds or more, the
The Signal Output Low Limit value is
displayed.
3
Enter "100" on the Display.
Reference
Move the digit to be entered using the right and left arrow buttons
( 0.1 digit
1 digit
At this time, the selected digit blinks.
Enter a numerical value ( 0, 1, 2, ..., 9 ) into the selected digit using the
upper and lower arrow buttons
value and
button decreases the value.
While pressing
changed.
To cancel the Signal Output Low Limit Setting, press
4
Press
to terminate the Signal Output Range Low Limit
Setting.
Reference
To change the Display for the Measurements Indication mode, press
.
Reference:
When Err6 is displayed, the value is out of the setting range.
300 mg/l ;
The Signal
100 digit ) .
10 digit
.
button increases a numeric
buttons, the values displayed are continuously
.
(2) Signal Output High Limit Setting
1
Press
Output High
"-Hi-" and "the setting value" comes on
alternatively.
2
Press
The Signal Output High Limit value is
displayed.
3
Enter "300" on the Display.
Pip
Reference
Move the digit to be entered using the right and left arrow buttons
( 0.1 digit
At this time, the selected digit blinks.
Enter a numerical value ( 0, 1, 2, ..., 9 ) into the selected digit using the
Pip Pip
upper and lower arrow buttons
value and
While pressing
changed.
To cancel the Signal Output High Limit Setting, press
4
Press
Limit Setting.
Reference
To change the Display for the Measurements Indication mode, press
Reference:
When Err6 is displayed, the value is out of the setting range.
and select the
The Signal
Limit.
for 2 seconds or more, the
100 digit ) .
1 digit
10 digit
.
button decreases the value.
buttons, the values displayed are continuously
to terminate the Signal Output Range High
.
Pip
Pip Pip
button increases a numeric
.