Diagnostics Mode - continued
Step-by-Step
Instructions for
dIAG Mode
Press the ZEROkey to
scroll through lists of
selections.
Press the TEST key to
make a selection.
If you want to skip a test,
press the ZERO key to
scroll to the next test.
dIAG will flash every
15 seconds during the
high resolution test to
remind you that you
are doing a test and
not seeing normal
weight readings.
5. Press the TEST key to perform
the RAM test. . .
6. Press the ZERO key twice to
scroll to the ROM test. . .
7. Press the TEST key to perform
the ROM test . . .
8. Press the ZERO key three times
to scroll to the I/O test . . . I/O is displayed. This stands for the
9. With a loopback connector in
place, press the TEST key to
perform the I/O test. . .
10. Press the ZERO key four times . . .
11. Press the TEST key to perform
this test. . .
12. Press the TEST key to stop
the test. . .
13. Press the ZERO key five times. . .
14. Press the TEST key to perform
this test. . .
PASS or FAIL is displayed briefly, then
dISP. If the test fails, contact your local
Weigh-Tronix distributor for service.
rO is displayed. This stands for the
ROM test.
PASS or FAIL is displayed briefly, then
dISP. If the test fails, contact your local
Weigh-Tronix distributor for service.
Input/Output test.
PASS or FAIL is displayed briefly, then
dISP. If the test fails, check your
connections and / or contact your
local Weigh-Tronix distributor for
service.
divA is displayed. This stands for the
high resolution test with AZT enabled.
The display shows the weight on the
scale at a resolution of 120,000 counts.
dISP is displayed.
divn is displayed. This stands for
the high resolution test without AZT
enabled.
The display shows the weight on the
scale at a resolution of 120,000
counts.