SECTION 9—SIEVE BEDS
SECTION 9—SIEVE BEDS

Replacing Sieve Beds

To prevent electrical shock, AWAYS disconnect concentrator from electrical outlet 
before servicing.
NOTE: For this procedure, refer to FIGURE 9.1 on page 45.
NOTE: ALWAYS replace sieve beds in pairs to ensure that both beds are in optimum condition.
1. Unplug unit.
2. Remove cabinet. Refer to Removing Cabinet on page 23.
3. Remove compressor inlet HEPA filter. Refer to Replacing the Compressor Inlet HEPA 
Filter on page 26.
4. Remove tie wraps and disconnect P.E valve tubing and check valve tubing from top 
of sieve beds.
5. Remove the four large tie‐wraps (two on each sieve bed).
6. Tilt sieve beds back and lift up. Rest sieve beds on rear of base.
7. Using a hose clamp tool, remove the clamp and reinforced tubing from bottom of both 
sieve beds and remove sieve beds.
DO NOT remove plastic caps from new sieve bed fittings until ready to install new beds. 
Severe sieve contamination can occur if uncapped beds are exposed to air.
8. Remove plastic caps from the top and bottom fittings of the new sieve beds.
9. Install new sieve beds reversing STEPS 3‐7.
NOTE: After replacing sieve beds, retiming is necessary. Refer to Timing on page 73.
10. Run unit and check for leaks. Refer to Leak Test on page 76.
11. Reinstall cabinet. Refer to Removing Cabinet on page 23.
Perfecto
2
DANGER
CAUTION
44
Part No 1148070