Invacare IRC5PO2 Service Manual page 22

Oxygen concentrator homefill compatible sensored and non-sensored
Hide thumbs Also See for IRC5PO2:
Table of Contents

Advertisement

SECTION 4—TROUBLESHOOTING
SYMPTOM
Unit Operating,
Internal Status Indicators:
RED: Three Flashes
GREEN: One Flash
Control Panel Indicators:
RED: Off
YELLOW: Flashing
GREEN: On
Low-Flow Alarm does not
activate on flows less than 0.5
L/min.
Perfecto
2
PROBABLE CAUSE
Internal repairs required.
System leak.
Defective check valves.
22
SOLUTION
Replace SensO
circuit board. Refer to
2
Replacing P.C. Board on page 58.
Repair leak in product tank, regulator,
tubing, fittings, or flow meter.
Replace check valves. Refer to Replacing
Check Valves on page 46.
Part No 1148070

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Irc5pPerfecto2 irc5po2Perfecto2 irc5p

Table of Contents