STATus:OPERation:CONDition?
Parameters
Comments
STATus:OPERation:ENABle < mask >
Parameters
Comments
Example
62 HP E1459A SCPI Command Reference
Returns the value of the Operation Status Condition Register as a signed 16 bit
integer.
None
•
The only bit in this register used by the HP E1459A is bit 9 (decimal weight
512) which contains the summary of the Operation Status Port register.
•
The Status Operation Condition register is not cleared by this command. It is
cleared only by executing the PSUMmary:EVENt command.
•
*RST clears all Status Operation Conditions.
•
*CLS does not affect the contents of the of the Status Operation Conditions.
•
The STATus:PRESet command does not affect the Status Operation
Conditions.
Sets the value of the OPERation Status Enable Register.
Parameter Name
Parameter Type
< mask >
numeric
•
<mask> determines which OPERation Status conditions are summed. See
Figure 3-1. The events detected in the Port Summary Status Register are
reported in bit 9 of the Operation Status Register which in turn is reported in
bit 7 of the Status Byte Register.
•
*RST and *CLS do not affect the value of the enable mask.
•
STATus:PRESet sets the value of the enable mask to 0.
STAT:OPER:ENAB 0xFFFF
Range of Values
-32768 to 32767 (0000
to FFFF
h
Enable all bits of the Operation Status
Enable Register
Default
)
0
h