Download Print this page

HP 8753ES Specifications page 8

Network analyzers 30 khz to 3 or 6 ghz
Hide thumbs Also See for 8753ES:

Advertisement

System performance summaries
HP 8753ES (75-ohm systems)
Type-F test ports
Configuration
Network analyzer
Calibration kit
Test-port cables
System dynamic range
30 kHz to 50 kHz
50 kHz to 300 kHz
300 kHz to 16 MHz
16 MHz to 3 GHz
Reflection uncertainty
Transmission uncertainty
1. Typical below 300 kHz.
2. Typical performance.
HP 8753E Option 075
HP 85039B
HP 11857B
68 dB
1
1
90 dB
96 dB
106 dB
Magnitude
Magnitude
Measurement port characteristics
Data is shown for a Type-F female reflection port and a
Type-F male transmission port.
Frequency range
2
Corrected
30 kHz-300 kHz
Directivity
45 dB
Source match
40 dB
Load match
45 dB
Reflection
±(0.060 dB
tracking
+0.02 dB/˚C)
Transmission
±(0.033 dB
tracking
+0.02 dB/˚C)
300 kHz-1.3 GHz 1.3 GHz-3 GHz
45 dB
40 dB
40 dB
30 dB
45 dB
40 dB
±(0.060 dB
±(0.024 dB
+0.01 dB/˚C)
+0.02 dB/˚C)
±(0.019 dB
±(0.057 dB
+0.01 dB/˚C)
+0.02 dB/˚C)
Phase
Phase