Download Print this page

HP 8753ES Specifications page 15

Network analyzers 30 khz to 3 or 6 ghz
Hide thumbs Also See for 8753ES:

Advertisement

HP 8753ET specifications
Test-port input characteristics (continued)
1
(CW time sweep only)
External source mode
Frequency range
R-channel input requirements
Power level
Spectral purity
Maximum spurious
input
Residual FM
Typical settling time
Frequency readout
2
accuracy
Input frequency accuracy requirement
Manual:
Accuracy
Magnitude characteristics
(10 Hz IF BW)
Dynamic accuracy
(300 kHz to 3 GHz)
(3 to 6 GHz)
0.001 dB/division
Display resolution
3
Marker resolution
Trace noise
(0 dBm at transmission port or +5 dBm at reflection port,
ratio measurement, 3 kHz BW)
300 kHz to 3 GHz
3 GHz to 6 GHz
1. See the HP 8753ET/ES descriptions and options for a functional description. Measurement accuracy is dependent on the
stability of the input signal.
2. Typical performance.
3. Marker resolution for magnitude, phase and delay is dependent upon measured value. Resolution is limited to five digits.
300 kHz to 6 GHz
2
0 to -25 dBm
<-30 dBc
<20 kHz
500 ms (automatic)
50 ms (manual)
0.1% (automatic)
2
-0.5 to 5 MHz
(See magnitude and phase
characteristics)
0.001 dB
< 0.006 dB rms
< 0.010 dB rms
Reference level
Range
(300 kHz to 3 GHz)
(3 to 6 GHz)
Resolution
2
Stability
300 kHz to 3 GHz
3 GHz to 6 GHz
Phase characteristics
(10 Hz IF BW)
Dynamic accuracy
Range
Display resolution
3
Marker resolution
Trace noise
(0 dBm at transmission port or +5 dBm at reflection
port, ratio measurement, 3 kHz BW)
300 kHz to 3 GHz
3 GHz to 6 GHz
Reference level
Range
Resolution
Stability
30 kHz to 3 GHz
3 GHz to 6 GHz
Polar characteristics
Range
Reference
±500 dB
0.001 dB
0.02 dB/ 1 ˚C
0.04 dB/ 1 ˚C
±180˚
0.01˚/ division
0.01˚
< 0.038˚ rms
< 0.070˚ rms
-180˚ to +180˚
0.01˚
0.05˚/ ˚C
0.20˚/ ˚C
-12
10 x 10
to 1000 units full
scale
±500 units
15

Advertisement

loading

This manual is also suitable for: