The System Maintenance screen is hidden from normal view. This screen, when active, allows the
user to select the treadmill type (Gait Trainer or RTM 600), perform speed and elevation calibra-
tion, and perform strain gauge calibration.
To access the System Maintenance screen:
1.
Touch <Utilities>.
Note: The next step requires pressing the hidden keypads in the upper right and left corners of the
displays touch screen.
2.
Touch the top right corner of the Utilities screen, the top left corner, and the top right corner
successively to access the <System Maintenance> prompt on the display.
3.
Touch <System Maintenance>. The System Maintenance screen should now be displayed.
elevation calibration
1.
From the System Maintenance screen, touch <Treadmill Diagnostic Display>.
2.
Touch M for the elevation until the elevation motor stops.
The reported elevation counts should be between 12-15. If not, loosen the
nut on the elevation pot and adjust the pot for a count between 12-15.
3.
Touch L until the treadmill is fully elevated. The counts should be between 121-125.
4.
Touch <Back> to return to the System Maintenance screen.
GAIT TRAINER UTILITIES
— 10-6 —