Plumbing Connection Details; Plumbing Installation - Invacare 6302G User Manual

Supine whirlpool tubs
Table of Contents

Advertisement

4 INSTALLATION
4.6

Plumbing Connection Details

Accommodate Reduced Pressure Zone (RPZ) if required.
4.7

Plumbing Installation

DANGER
Be certain all plumbing work is performed by a qualified plumber and is in compliance with local codes. Check local codes regarding
maximum water temperature allowed at hot water outlets for facility type.
WARNINGS
Be certain all plumbing work is performed by a qualified plumber and is in compliance with local codes.
Some state and local plumbing codes require the installation of a "reduced pressure zone" (RPZ) assembly on both the incoming hot and
cold supply lines to prevent back flow contamination into the potable water system. Invacare recommends you check local plumbing
code requirements to determine if RPZs are required on your tub installation. RPZs are NOT supplied by Invacare and are to be
provided by the customer if required.
Invacare® Supine Whirlpool Tubs
3/4 in. Ball Valve
minimum of 6 in
above the floor
FIGURE 4.1 Plumbing Connection Details
MUST be a
Centerline of Tub
Tub
Front of Tub
2 in. PVC stub
18
Wall
Drain
38 in.
(965 mm)
5 in.
(127 mm)
Back of Tub
Elbow
Drain
Part No. 1167456

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

6302gadh

Table of Contents