Configuration And Wiring - Bosch CRS-RM-905FI User Manual

Wall transmitter
Table of Contents

Advertisement

NurseCall N46 Wall Transmitter
5.2.8

Configuration and Wiring

CR 2450
battery
Reset
Connector J4:
External Call for Help
trigger (pear push button)
(o)
Solder bridges B1 and B2 are located on the bottom side of the PCB:
Fig. 5.2 Configuration and Wiring
NOTICE!
i
Setting up one or more inputs to "NC" position results to higher power consumption. This con-
figuration should be used with external power supply or will reduce the battery life.
Bosch Security Systems
N
NC
N
NC
J
J5
1 2 3 4 5 6
N
NC
J
J4
5
4
3
2
1
NC-B2-NO
NC-B1-NO
Solder Bridges (if soldered):
B8 Call for Assistance only by
B7
Call for Assistance from terminals J6[4-5]
B6 No
Staff Presence
B5 No Daily Message
B4 No calling message repetition
B3 Technical Alarm message repetition
(o)
B1 REL 1 Call for Help
(o)
B2 REL 2 Staff Presence
Terminal J5
1 External 12-24 VDC power supply
2 GND (0 V)
3-4 REL 1 Call for Help
5-6 REL 2 Staff Presence
Connector J7:
Presence / Staff identification (optional)
Terminal J6:
4-5 Reserve Call or Call for Assistance
3-4 Technical Alarm
1-2 Call for Help
J6
J1 External Call for Help trigger
input polarity
J2 Polarity of Technical Alarm input
J3 Polarity of
User Manual
Installation | en
Staff Presence
processing
polarity
polarity
terminal J6[4-5]
input
970.030 | v1.1 | 2009.09
17

Advertisement

Table of Contents
loading

Table of Contents