Page 1
PS-500 SCALE Operation Manual Contents subject to change without notice Version 1.0 2006-01-10 Declarations of compliance...
Page 2
Wa r n i n g s Electrical installation United States This equipment has been tested and found to comply with the limits for a Class A For your protection, all mains (11 0V or 230V) equipment used digital device, pursuant to Part 15 of the FCC Rules. These limits are designed to where damp or wet conditions may occur, must be supplied from a provide reasonable protection against harmful interference when the equipment correctly fused source and protected by an approved ground fault protection...
Page 3
Operation Manual CONTENT Thank you for purchasing the indicator. Please read all operating instructions carefully before use and keep the following points in mind: * Avoid lengthy exposure to extreme heat or cold, your scale works best when operated at normal room temperature. Always allow the scale to acclimate to a normal room Ⅰ.
Page 4
standard SCP-01) Ⅱ. KEY FUNCTION: 5. Output data: gross weight, net weight, tare weight, indicator displaying weight, weighing unit etc. FACEPLATE: ·OPERATION INTERFACE: 1. Display: 0.65” (17mm) 7-segment LCD, 5 digits 2. Keyboard: 4-key push button ·POWER: 1. Alkaline Batteries: 4 x “AAA” size cells When all displayed segments of LCD flashed, this indicates the batteries are low below 4.9V and you’d better to replace batteries;...
Page 5
HOLD c. If this key is set for both HOLD and PRINT function (P2=2), press this key to PRINT output the data according to P4 setting. Hold these two buttons to show firmware version; A/D code or input UNIT working voltage of indicator. Choose weighing units among kg-lb-lb: oz UNIT Note: The weighing units that can be used are restricted by display...
Page 6
3.4 In displaying A/D code or input voltage mode: of FS weight, and the unit is same as P10 setting) for calibration. Move away any weight on scale. When normal weighing mode, press and hold down TARE and ON/OFF/ZERO buttons to enter calibration mode. When the indicator shows”...
Page 7
zero-point. Move away any weight on the scale. Press TARE button to confirm, store all parameters in EEPROM. And then it will auto-reset and display all or press ON/OFF/ZERO to exit this mode. segments of LCD, full capacity… like power on again. If there’s an error occurred When ‘CAL.P1’...
Page 8
is error; adjust potentiometer’s position to make the ADC data will be positive HOLD/PRINT key to shift the flashed position. Press TARE key to confirm and value and larger than (Zp%+ Zk%) x NFS. Normally the indicator is save the set data and enter next setting. Press ON/OFF/ZERO key to exit this factory-calibrated, and end users do not need this operation.
Page 9
displayed weight reading and weight unit, and not receive any data from x=2: 7O1 7 digits, 1 odd, 1 start bit, 1stop bit, other equipment. The output format is as below: <LF>< reading, minus, decimal point, weight unit><CR><EXT> 5.7) P7.xy: calibration resolution x=2: Press PRINT/HOLD key, the scale will output the data of stable Table2: current gross, tare, net weight reading and weight unit.
Page 10
5.10) P10.x: Calibration unit 1 lb 0.5 kg 1 lb Not available x=0: kg as calibration unit 10 lb 5 kg 10 lb Not available x=1: lb as calibration unit 0.0002 lb 0.0001 kg 0.0002 lb Not available 0.002 lb 0.001 kg 0.002 lb Not available...
Page 11
x=7: No limitation 5.17) P17.x: Data filter intensity 5.13) P13.x: Zero range for ZERO button after switch on x=0 very weak x=0: the range is set to calibration zero point (CAL.P0) +1%FS x=1 weak x=1: the range is set to calibration zero point (CAL.P0) +2%FS x=2 middle x=2: the range is set to calibration zero point (CAL.P0) +5%FS x=3 strong...
Page 12
GND 5 --------------- 5 ------------------------ 5 --------------5 GND <ETX> End of Text character (hex 03) DSR 4 --------------- 4 ------------------------- 4 --------------4 DTR <SP> Space (hex 20H) DTR 6 --------------- 6 ------------------------- 6 --------------6 DSR H1H2H3 Three status bytes CTS 7 --------------- 7 -------------------------- 7 --------------7 RTS <p>...
Page 13
Response: Zero function is activated and it returns to current scale status. just like 11= not define 1= at zero point 1= over capacity pressing ZERO/ON/OFF button: 0= gross weight <LF> H1H2H3 <CR><ETX> always 0 always 0 1= net weight If ZERO function cannot be activated, it will return to current scale status.
Page 14
Table7: (Capacity unit is kg or lb) 3000 0.30000 3.000 30.00 300.0 3000 30000 Resolution Division set by P8 and P9( P8=1) 3500 0.3500 3.500 35.00 350.0 3500 35000 set by P7 0.0002 0.002 0.02 4000 0.4000 4.000 40.00 400.0 4000 40000 0.1000...
Page 16
EEP.E1---------The parameters are not same with backup data; EEP.E2---------The setting parameter(s) is not in normal range;. 35000 17.5000 175.000 1750.00 17500.0 17500 1750000 CAL-Px --------scale’s calibration point; CAL.Er ---------there is an error in calibration Hold ----------hold function is active. 40000 20.0000 200.000 2000.00...
Page 17
Setup mode/Calibration mode/Displaying A/D code Exit and back to normal work mode or input voltage mode ON/OFF/ ZERO Normal weighing mode Go to Show A/D code or input working Ⅹ.Key Definition summary: (more than 3s) voltage of indicator mode HOLD /PRINT MODE DEFINITION...
Page 18
0=no RS232 function; 1=output display data when Zero range for ZERO button: PRINT pressed; 2=output gross, tare and net 0= Power-on zero-point +1%FS; weight when PRINT pressed; 3=continuously 1= Power-on zero-point +2FS; 0,1,2,3,4, 2= Power-on zero-point +5%FS; output display data; 4= continuously output gross, 0,1,2,3,4, P4.x 5,6,7...
Page 19
B66 2TE Tel: +44 (0) 870 444 6132 Fa x : +44 (0) 870 010 2241 Email: sales@salterbrecknell.co.uk Web site: www.averyweigh-tronix.com Salter Brecknell Weighing Products USA 1000 Armstrong Drive Fairmont MN 56031 Toll Free: 800-637-0529 Phone: 507-238-8702 Fax: 507-238-8271 Email: sales@salterbrecknell.com Web site: www.salterbrecknell.com...