Wireless Blood Pressure Monitor (BP5) OWNER’S MANUAL Table of Contents INTRODUCTION PACKAGE CONTENTS INTENDED USE BLOOD PRESSURE CLASSIFICATION FOR ADULTS CONTRAINDICATION PARTS AND DISPLAY INDICATORS SET UP REQUIREMENTS SET UP PROCEDURES MEASUREMENT PROCEDURES SPECIFICATIONS GENERAL SAFETY AND PRECAUTIONS BATTERY HANDLING AND USAGE...
INTRODUCTION Thank you for selecting the iHealth Wireless Blood Pressure Monitor. The iHealth Wireless Blood Pressure Monitor is a fully automatic arm cuff blood pressure monitor that uses the oscillometric principle to measure your blood pressure and pulse rate. The monitor works with your mobile device to test, track and share vital blood pressure data.
BLOOD PRESSURE CLASSIFICATION FOR ADULTS The World Health Organization (WHO) has created the following guide for assessing high blood pressure (without regard to age or gender). Please note that other factors (e.g. diabetes, obesity, smoking, etc.) also need to be considered. Consult with your physician for accurate assessment. Classification of blood pressure for adults Systolic BLOOD PRESSURE...
Status indicator START/STOP button Cuff SET UP REQUIREMENTS The iHealth Wireless Blood Pressure Monitor is designed to be used with the following iPod touch, iPhone and iPad models: iPod touch (5th generation) iPod touch (4th generation) iPod touch (3rd generation)
When using the monitor for the first time, it may take up to 30 seconds for your iOS device to detect the Bluetooth signal. • Each subsequent time you use the monitor, "Not Connected" will be displayed next to "BP5 xxxxxx" in the Bluetooth Menu.
• Please repeat these steps when you switch to another iOS device with the monitor. Monitor Status Bluetooth Indicator Waiting to connect Flashing blue light Connected and measuring Steady blue light Measurement completed and ready to disconnect Gradually extinguishing light MEASUREMENT PROCEDURES Blood pressure can be affected by the position of the cuff and your physiologic condition.
Page 7
Apply the Cuff a. Pull the cuff end through the metal loop, positioning it outward 1/2” (1-2cm (away from your body). b. Place a bare arm through the cuff and position the cuff 1/2” (1-2cm) above the elbow joint. c. Tighten the cuff by pulling it towards your body, securing it closed with the Velcro fastener.
Page 8
5. Keep the cuff clean. Cleaning the cuff after every 200 times of usage is recommended.If the cuff becomes dirty, clean it with a moistened cloth. Do not rinse the monitor or cuff with running water. During measurement, press the "START/STOP" button to stop measurement. Press the "START/STOP"...
14. Environmental humidity for operation: ≤90%RH 15. Environmental temperature for storage and transport: -20°C~55°C(-4°F~131°F) 16. Environmental humidity for storage and transport: ≤90%RH 17. Environmental pressure: 80kPa-105kPa 18. Battery life: more than 80 measurements on a full charge 19. The blood pressure measurement system includes accessories: pump, valve, cuff, and sensor. Note: These specifications are subject to change without notice.
Page 11
automatically. It is recommended that the iOS device be set in Airplane mode during measure- ment to avoid strong magnetism interference. 6. If Irregular Heartbeat (IHB) is detected during the measurement procedure, the IHB symbol will be displayed. Under this condition, the Wireless Blood Pressure Monitor can keep functioning, but the results may be inaccurate.
12. A medical AC adapter with an output of DC 5.0V and complies with IEC 60601-1/UL 60601-1 and IEC 60601-1-2/EN 60601-1-2 is suitable for this monitor, such as ASP5-05010002JU (input: 100-240V, 50/60Hz, 200mA; output: DC 5.0V, 1.0A). Please note that the monitor jack size is USB mini B.
Page 13
Monitor Status Status Indicator Charging Flashing green light Fully charged Steady green light Low battery Flashing red light (for a few seconds) Abnormal state Steady red light Do not change the battery. If the battery can no longer be charged, please contact Customer Service.
TROUBLESHOOTING PROBLEM POSSIBLE CAUSE SOLUTION Low Battery Battery is less than 25% Charge the battery Blood pressure is outside of measurement range Retest, make sure your blood pressure is within measurement range Arm or monitor was moved during test Retest, make sure not to move your arm or the monitor Display reads The cuff does not inflate properly or pressure Review the cuff application instructions and retest...
Avoid high temperature and direct sunlight. Do not immerse the monitor in water as this will result in damage to the monitor. Do not attempt to disassemble this monitor. Battery replacement should only be performed by a qualified iHealth technician. To do...
WARRANTY INFORMATION The iHealth Wireless Blood Pressure Monitor is warranted to be free from defects in materials and workmanship within one year from the date of purchase when used in accordance with the provided instructions. The warranty extends only to the end user. We will, at our option, repair or replace without charge the iHealth Wireless Blood Pressure Monitor covered by the warranty.
Page 17
Apple Inc., registered in the U.S. and other countries. The Bluetooth word mark and logos are registered trademarks owned by Bluetooth SIG, Inc. ® and any use of such marks by iHealth Lab Inc. is under license. Other trademarks and trade names are those of their respective owners.
(2) this device must accept any interference received, including interference that may cause undesired operation. Changes or modifications not expressly approved by iHealth Lab Inc. would void the user’s authority to operate the product. Note: This product has been tested and found to comply with the limits for a Class B digital device, pursuant to Part 15 of the FCC Rules.
to radio or television reception, which can be determined by turning the equipment off and on, the user is encouraged to try to correct the interference by one or more of the following measures: —Reorient or relocate the receiving antenna. —Increase the separation between the equipment and receiver.
ANSI/AAMI SP-10:2002+A1: 2003+A2:2006; AAMI/ANSI 80601-2-30: 2009/IEC 80601-2-30: 2009+Cor.2010/EN 80601-2-30:2010(Medical electrical equipment –Part 2-30: Particular requirements for the basic safety and essential perfor- mance of automated non-invasive sphygmomanometers). ELECTROMAGNETIC COMPATIBILITY INFORMATION Table 1 For all ME EQUIPMENT and ME SYSTEMS Guidance and manufacture’s declaration - electromagnetic emissions The Wireless Blood Pressure Monitor is intended for use in the electromagnetic environment specified below.
Page 21
Table 2 For all ME EQUIPMENT and ME SYSTEMS Guidance and manufacturer’s declaration - electromagnetic immunity The Wireless Blood Pressure Monitor is intended for use in the electromagnetic environment specified below. The customer or the user of the Wireless Blood Pressure Monitor should assure that it is used in such an environment. IMMUNITY test IEC 60601test level Compliance level...
Page 22
Power frequency 3 A/m 3 A/m Power frequency magnetic fields should be at levels (50/60 Hz) characteristic of a typical location in a typical magnetic field commercial or hospital environment. IEC 61000-4-8 NOTE: U is the a.c. mains voltage prior to application of the test level. Table 3 For ME EQUIPMENT and ME SYSTEMS that are not LIFE-SUPPORTING Guidance and manufacturer’s declaration - electromagnetic immunity...
Page 23
distance in meters (m). Field strengths from fixed RF transmitters, as determined by an electromagnetic site survey, should be less than the compliance level in each frequency range. Interference may occur in the vicinity of equipment marked with the following symbol: NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
Page 24
Wireless Blood Pressure Monitor Tensiomètre sans fil Misuratore di pressione wireless Monitor de presión arterial inalámbrico Drahtloses blutdruckmessgerät OWNER’S MANUAL MANUEL DE L'UTILISATEUR MANUALE D’ISTRUZIONI MANUAL DEL PROPIETARIO BEDIENUNGSANLEITUNG...
Page 25
Table 4 For ME EQUIPMENT and ME SYSTEMS that are not LIFE-SUPPORTING Recommended separation distances between portable and mobile RF communications equipment and the Wireless Blood Pressure Monitor The Wireless Blood Pressure Monitor is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled.