    Wireless Blood Pressure Monitor (BP5) OWNER’S MANUAL Table of Contents INTRODUCTION PACKAGE CONTENTS INTENDED USE BLOOD PRESSURE CLASSIFICATION FOR ADULTS CONTRAINDICATION PARTS AND DISPLAY INDICATORS SET UP REQUIREMENTS SET UP PROCEDURES MEASUREMENT PROCEDURES SPECIFICATIONS GENERAL SAFETY AND PRECAUTIONS BATTERY HANDLING AND USAGE...
  • Page 2: Introduction

    INTRODUCTION Thank you for selecting the iHealth Wireless Blood Pressure Monitor. The iHealth Wireless Blood Pressure Monitor is a fully automatic arm cuff blood pressure monitor that uses the oscillometric principle to measure your blood pressure and pulse rate. The monitor works with your mobile device to test, track and share vital blood pressure data.
  • Page 3: Blood Pressure Classification For Adults

    BLOOD PRESSURE CLASSIFICATION FOR ADULTS The World Health Organization (WHO) has created the following guide for assessing high blood pressure (without regard to age or gender). Please note that other factors (e.g. diabetes, obesity, smoking, etc.) also need to be considered. Consult with your physician for accurate assessment. Classification of blood pressure for adults Systolic BLOOD PRESSURE...
  • Page 4: Set Up Requirements

    Status indicator START/STOP button Cuff SET UP REQUIREMENTS The iHealth Wireless Blood Pressure Monitor is designed to be used with the following iPod touch, iPhone and iPad models: iPod touch (5th generation) iPod touch (4th generation) iPod touch (3rd generation)
  • Page 5: Set Up Procedures

    SET UP PROCEDURES Download the Free iHealth MyVitals App Prior to first use, download and install "iHealth MyVitals" from the App Store. Account Set Up and Registration Follow the on-screen instructions to register and set up your personal account. Charge Battery Before First Use Connect the monitor to a USB port using the charging cable provided until the green indicator light stabilizes.
  • Page 6: Measurement Procedures

    ‹ 7SLHZL YLWLH[ [OLZL Z[LWZ ^OLU `V\ Z^P[JO [V HUV[OLY P6: KL]PJL ^P[O [OL TVUP[VY Monitor Status Bluetooth Indicator Waiting to connect Flashing blue light Connected and measuring Steady blue light Measurement completed and ready to disconnect Gradually extinguishing light MEASUREMENT PROCEDURES Blood pressure can be affected by the position of the cuff and your physiologic condition.
  • Page 7 Apply the Cuff a. Pull the cuff end through the metal loop, positioning it outward 1/2” (1-2cm) (away from your body). b. Place a bare arm through the cuff and position the cuff ” JT) above the elbow joint. c. Tighten the cuff by pulling it towards your body, securing it closed with the Velcro fastener.
  • Page 8 5. Keep the cuff clean. Cleaning the cuff after every 200 times of usage is recommended.If the cuff becomes dirty, clean it with a moistened cloth. Do not rinse the monitor or cuff with running water. During measurement, press the "START/STOP" button to stop measurement. Press the "START/STOP"...
  • Page 9: Specifications

    SPECIFICATIONS  7YVK\J[ UHTL! >PYLSLZZ )SVVK 7YLZZ\YL 4VUP[VY 2. Model: BP5 3. Classification: Internally powered, Type BF applied part, IPX0, No AP or APG, Continuous operation 4. Machine size: approx. 5.7” x 2.3” _ ”TT í TT í TT 5. Cuff circumference: 8.6””JTJT ” ”(42cm-48cm) (XL size sold separately)  >LPNO[! HWWYV_ Va N PUJS\KPUN J\MM"...
  • Page 10: General Safety And Precautions

     ,U]PYVUTLU[HS O\TPKP[` MVY VWLYH[PVU! ˜  9/  ,U]PYVUTLU[HS [LTWLYH[\YL MVY Z[VYHNL HUK [YHUZWVY[! °C~55°C(-4°F~°F)  ,U]PYVUTLU[HS O\TPKP[` MVY Z[VYHNL HUK [YHUZWVY[! ˜  9/  ,U]PYVUTLU[HS WYLZZ\YL! R7HR7H  )H[[LY` SPML! TVYL [OHU  TLHZ\YLTLU[Z VU H M\SS JOHYNL   ;OL ISVVK WYLZZ\YL TLHZ\YLTLU[ Z`Z[LT PUJS\KLZ HJJLZZVYPLZ! pump, valve, cuff, and sensor. Note: These specifications are subject to change without notice.
  • Page 11 automatically. It is recommended that the iOS device be set in Airplane mode during measure- ment to avoid strong magnetism interference. 6. If Irregular Heartbeat (IHB) is detected during the measurement procedure, the IHB symbol will be displayed. Under this condition, the Wireless Blood Pressure Monitor can keep functioning, but the results may be inaccurate.
  • Page 12: Battery Handling And Usage

     ( TLKPJHS (* HKHW[LY ^P[O HU V\[W\[ VM +* = HUK JVTWSPLZ ^P[O 0,* <3  HUK 0,* ,5  PZ Z\P[HISL MVY [OPZ TVUP[VY Z\JO HZ (:71< PUW\[! = /a T(" V\[W\[! +* = ( 7SLHZL UV[L [OH[ [OL TVUP[VY QHJR size is USB mini B.
  • Page 13 Monitor Status Status Indicator Charging Flashing green light Fully charged Steady green light Low battery Flashing red light (for a few seconds) Abnormal state Steady red light Do not change the battery. If the battery can no longer be charged, please contact Customer Service.
  • Page 14: Troubleshooting

    TROUBLESHOOTING PROBLEM POSSIBLE CAUSE SOLUTION Low Battery Battery is less than 25% Charge the battery Blood pressure is outside of measurement range Retest, make sure your blood pressure is within measurement range Arm or monitor was moved during test Retest, make sure not to move your arm or the monitor Display reads The cuff does not inflate properly or pressure Review the cuff application instructions and retest...
  • Page 15: Care And Maintenance

    Avoid high temperature and direct sunlight. Do not immerse the monitor in water as this will result in damage to the monitor. Do not attempt to disassemble this monitor. Battery replacement should only be performed by a qualified iHealth technician. To do...
  • Page 16: Warranty Information

    WARRANTY INFORMATION The iHealth Wireless Blood Pressure Monitor is warranted to be free from defects in materials and workmanship within one year from the date of purchase when used in accordance with the provided instructions. The warranty extends only to the end user. We will, at our option, repair or replace without charge the iHealth Wireless Blood Pressure Monitor covered by the warranty.
  • Page 17 Apple Inc., registered in the U.S. and other countries. ® The Bluetooth word mark and logos are registered trademarks owned by Bluetooth SIG, Inc. and any use of such marks by iHealth Lab Inc. is under license. Other trademarks and trade names are those of their respective owners.
  • Page 18: Important Information Required By The Fcc

    Manufactured for iHealth Lab Inc.  5 :OVYLSPUL )S]K 4V\U[HPU =PL^ *(  <:(  www.ihealthlabs.com Lotus Global Co., Ltd.  (SL_HUKYH 9VHK 3VUKVU <2 5> +7 ;LS!  -H_!   ANDON HEALTH CO., LTD. 5V  1PUWPUN :[YLL[ @H(U 9VHK 5HURHP +PZ[YPJ[ ;PHUQPU   *OPUH ;LS! ...
  • Page 19: Other Standards And Compliances

    to radio or television reception, which can be determined by turning the equipment off and on, the user is encouraged to try to correct the interference by one or more of the following measures: —Reorient or relocate the receiving antenna. —Increase the separation between the equipment and receiver.
  • Page 20: Electromagnetic Compatibility Information

    (5:0((40 :7!(! (!" ((40(5:0 !  0,* !  *VY,5 !4LKPJHS electrical equipment –Part 2-30: Particular requirements for the basic safety and essential perfor- mance of automated non-invasive sphygmomanometers). ELECTROMAGNETIC COMPATIBILITY INFORMATION Table 1 For all ME EQUIPMENT and ME SYSTEMS Guidance and manufacture’s declaration - electromagnetic emissions The Wireless Blood Pressure Monitor is intended for use in the electromagnetic environment specified below.
  • Page 21 Table 2 For all ME EQUIPMENT and ME SYSTEMS Guidance and manufacturer’s declaration - electromagnetic immunity The Wireless Blood Pressure Monitor is intended for use in the electromagnetic environment specified below. The customer or the user of the Wireless Blood Pressure Monitor should assure that it is used in such an environment. IMMUNITY test IEC 60601test level Compliance level...
  • Page 22 Power frequency 3 A/m 3 A/m Power frequency magnetic fields should be at levels (50/60 Hz) characteristic of a typical location in a typical magnetic field commercial or hospital environment. 0,*  NOTE: U is the a.c. mains voltage prior to application of the test level. Table 3 For ME EQUIPMENT and ME SYSTEMS that are not LIFE-SUPPORTING Guidance and manufacturer’s declaration - electromagnetic immunity...
  • Page 23 distance in meters (m). Field strengths from fixed RF transmitters, as determined by an electromagnetic site survey, should be less than the compliance level in each frequency range. Interference may occur in the vicinity of equipment marked with the following symbol: 56;,  ([  4/a HUK  4/a [OL OPNOLY MYLX\LUJ` YHUNL HWWSPLZ NOTE 2 These guidelines may not apply in all situations.
  • Page 24 Table 4 For ME EQUIPMENT and ME SYSTEMS that are not LIFE-SUPPORTING Recommended separation distances between portable and mobile RF communications equipment and the Wireless Blood Pressure Monitor The Wireless Blood Pressure Monitor is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled.