U S E R M A N U A L xygen OnitOr Model: PM5900 SAVE THESE INSTRUCTIONS CAUTION Federal (USA) law restricts this device to sale by or on the order of a physician. 300 Held Drive Tel: (+001) 610-262-6090...
CONTENTS RECEIVING/INSPECTION ............... 2 INTENDED USE ..................2 OPERATOR PROFILE ................2 READ ALL INSTRUCTIONS BEFORE USING ........2 SAFETY INFORMATION - WARNINGS AND CAUTIONS ....... 2 SPECIFICATIONS ..................4 SPECIFICATIONS ARE SUBJECT TO CHANGE WITHOUT PRIOR NOTICE ....................4 COMPONENT DESCRIPTION ..............
The oxygen monitor is to be used by trained healthcare professionals under the supervision, or on the order, of a physician in a hospital (or other clinical setting). The Precision Medical, Inc. Oxygen Monitor is not intended for transport use. This device is not an oxygen supply source.
Page 4
• Always follow ANSI and CGA standards for Medical Gas Products, Flowmeters, and Oxygen Handling. • The Oxygen Monitor should be serviced by a qualified hospital/ dealer service technician, or by Precision Medical, Inc. • DO NOT obstruct the alarm grill on the back of the Oxygen Monitor.
COMPONENT DESCRIPTION CAUTION Missing or illegible labels must be replaced, contact Precision Medical, Inc. OXYGEN MONITOR IDENTIFICATION 1. High Alarm Set Key 2. ↑ Up & 7. High Alarm LED ↓ Down Keys 3. Low Alarm 8. LCD Display Set Key 9.
Page 7
NO. ITEM NAME DESCRIPTION High Alarm Pressing the key when the keypad Set Key is unlocked, the high alarm set point will flash. The high alarm setting can be adjusted with the UP/DOWN keys. When the high alarm value is changed, pressing the key will save the setting.
Page 8
ITEM NAME DESCRIPTION Power Key key turns the Oxygen Monitor ON and OFF. The LOCK/UNLOCK key must be pressed to unlock the Oxygen Monitor, before being powered OFF. Lock / Unlock Pressing the key will “Unlock ” the keypad, so changes in the stored settings can be made.
OXYGEN SENSOR COMPONENT IDENTIFICATION 23. Coiled Cable 24. Locking Ring 25. Oxygen Sensor 26. Diveter Fitting 27. Tee Adaptor NO. ITEM NAME DESCRIPTION 23 Coiled Cable The Coiled Cable allows the Sensor to be positioned up to 10 ft from the side with Male Plugs of the Oxygen Monitor.
BATTERY INSTALLATION CAUTION Use ONLY AA Alkaline Batteries. 1. Press down the center, top edge of the battery cover and slide down. 2 . Remove old batteries, if applicable. 3. Install 4 AA Alkaline Batteries, and follow the diagram stamped in the bottom of the battery compartment. 4.
NEW Sensor. Leaking or used Sensors should be handled and disposed of in accordance with local regulations. An MSDS is available from Precision Medical, Inc. • If the Oxygen Sensor is used in breathing circuits, the Diverter must be attached to the Sensor and must be used with the Tee Adapter.
• Calibrate the Oxygen Monitor before each use, and when replacing the Oxygen Sensor or the batteries. • The Precision Medical Oxygen Monitor can only be calibrated accurately using 100% Oxygen or 20.9% Oxygen (Room Air). Using any other concentration will result in a inaccurate readings.
6. Press the CAL key. The display will show “CAL @ Air” or “CAL @ 100” oxygen. When “DONE” appears, the calibration is complete. The Oxygen Concentration will be displayed in a gas percent value. Press LOCK/UNLOCK to save the calibration data.
The following recommendation is provided to reduce the chances of pressure causing false readings Calibrate the Precision Medical Oxygen Monitor using 100% Oxygen or room air at the same pressure and flow as the gas to be monitored. EFFECTS OF ANESTHETIC GASES:...
ALARMS The Precision Medical Oxygen Monitor will store the HIGH/LOW alarm settings in memory after the Oxygen Monitor is turned “OFF”. The Oxygen Monitor is designed to prevent crossing of the HIGH/ LOW alarm settings. The LOW alarm cannot be set above the HIGH alarm and the High alarm cannot be set below than the LOW alarm.
ALARM SITUATION: During a HIGH or LOW alarm. 1. Triple pulse alarm sounds. 2. LED will flash. 3. Alarm Low/High set point will flash. 4. UP or DOWN Arrow will turn dark. Pressing the Alarm Silent key will deactivate the audible alarm for 120 seconds.
(RGA) number, contact Precision Medical, Inc. All returns must be packaged in sealed containers to prevent damage. Precision Medical, Inc. will not be responsible for goods damaged in transit. Refer to Precision Medical, Inc. Return Policy available on the Internet, www.precisionmedical.com. DISPOSAL INSTRUCTIONS The Oxygen Sensor contains lead.
Page 19
Oxygen Condensation to changes Sensor 2. Replace with New in oxygen 2. Non functioning Precision Medical concentration Oxygen Sensor Oxygen Sensor E. Triple Pulse 1. Oxygen readings 1. Adjust the High/Low Alarm and are outside the...
(at Precision Medical, Inc.’s option) at its own expense. Precision Medical, Inc. warrants the 504877 Oxygen Sensor included with the PM5900 Oxygen Monitor to be free from defects in material and workmanship for a period of sixteen (16) months, from date of shipment. Should any failure to conform to this warranty appear within the applicable period, Precision Medical, Inc.
Page 23
Precision Medical, Inc. 300 Held Drive Northampton, PA 18067 USA Toll Free Tel: 800-272-7285 Toll Free Fax: 800-353-1240 Tel:(+001) 610-262-6090 Fax:(+001) 610-262-6080 www.precisionmedical.com ISO 13485 Certified...